Phenoxyethanol
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
C8H10O2 138.16
2-Phenoxyethanol;
2-Phenoxyethyl alcohol;
Ethylene glycol, 2-monophenyl ether CAS RN®: 122-99-6.
1 DEFINITION
Phenoxyethanol contains NLT 98.0% and NMT 102.0% of phenoxyethanol (C8H10O2).
2 IDENTIFICATION
Change to read:
A. Spectroscopic Identification Tests 〈197〉, Infrared Spectroscopy: 197F. (CN 1-May-2020) On an undried specimen
3 ASSAY
3.1 Procedure
Phenol solution, Standard solution, and Chromatographic system: Prepare as directed in the test for Organic Impurities.
Sample stock solution: 5 mg/mL of Phenoxyethanol in isopropyl alcohol
Sample solution: Transfer 500 μL of the Sample stock solution to a vial, add 1000 μL of isopropyl alcohol, crimp the vial, and mix on a vortex mixer for 15 s.
3.2 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of phenoxyethanol (C8H10O2) in the portion of Phenoxyethanol taken:
Result = (rU/rS ) × (CS/CU) × 100
rU = peak response from the Sample solution
rS = peak response from the Standard solution
CS = concentration of USP Phenoxyethanol RS in the Standard solution (mg/mL)
CU = concentration of Phenoxyethanol in the Sample solution (mg/mL)
Acceptance criteria: 98.0%–102.0%
4 IMPURITIES
Organic Impurities
Phenol solution: 0.25 mg/mL of phenol in isopropyl alcohol
Standard stock solution: 5 mg/mL of USP Phenoxyethanol RS in the Phenol solution
Standard solution: Transfer 500 μL of the Standard stock solution to a vial, add 1000 μL of isopropyl alcohol, crimp the vial, and mix on a vortex mixer for 15 s.
Sample solution: Transfer 500 μL of Phenoxyethanol to a tared vial, and determine the weight of Phenoxyethanol taken. Add 1000 μL of isopropyl alcohol, crimp the vial, and mix on a vortex mixer for 15 s.
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: GC
Detector: Flame ionization
Column: 0.32-mm × 10-m capillary coated with a 5-μm film of stationary phase G27
Temperatures
Injection port: 300°
Detector: 300°
Column: See Table 1.
Table 1
Initial Temperature (°) | Temperature Ramp (°/min) | Final Temperature (°) | Hold Time at Final Temperature (min) |
| 80 | 8 | 260 | 10 |
Carrier gas: Helium
Injection volume: 1 μL
Injection type: Split injection mode
Split flow rate: 44 mL/min
System suitability
Sample: Standard solution
Suitability requirements
Resolution: NLT 10 between the phenol and phenoxyethanol peaks
Relative standard deviation: NMT 2.0% for the phenoxyethanol peak
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of total impurities in the portion of Phenoxyethanol taken:
Result = (rU /rS ) × (CS /CU ) × 100
rU = peak response of all additional peak areas in the Sample solution, excluding the main peak, the solvent peak, and the phenol peak
rS = peak response of phenoxyethanol from the Standard solution
CS = concentration of phenoxyethanol in the Standard solution (mg/mL)
CU = concentration of the Sample solution (mg/mL)
Acceptance criteria: NMT 1.0%
Limit of Phenol
Phenol solution, Standard solution, Sample solution, and Chromatographic system: Proceed as directed in the test for Organic Impurities.
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of phenol in the portion of Phenoxyethanol taken:
Result = (rU /rS ) × (CS /CU ) × 100
rU = peak response of phenol from the Sample solution
rS = peak response of phenol from the Standard solution
CS = concentration of phenol in the Standard solution (mg/mL)
CU = concentration of the Sample solution (mg/mL)
Acceptance criteria: NMT 0.1%
5 SPECIFIC TESTS
Specific Gravity 〈841〉: 1.105–1.110 at 20°
6 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in tight containers, and store in a cool, dry place, protected from light.
USP Reference Standards 〈11〉
USP Phenoxyethanol RS