Oxygen 93 Percent
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Oxygen 93 Percent is Oxygen produced from air by the molecular sieve process. It contains NLT 90.0% and NMT 96.0% by volume of oxygen (O2), the remainder consisting mostly of argon and nitrogen.
2 IDENTIFICATION
A. The paramagnetic signal exhibited by the Sample gas in the Assay confirms the presence of oxygen.
B. The Sample gas in the Assay meets the Acceptance criteria.
3 ASSAY
Procedure
The certified standards called for in the following test are listed in Reagents, Indicators, and Solutions.
Zero gas: Nitrogen certified standard
Span gas: 93% Oxygen certified standard
Sample gas: Oxygen 93 Percent under test
Mode: Paramagnetic oxygen measurement (see Medical Gases Assay 〈415〉)
Analysis: Determine the concentration of oxygen in percentage by volume in Oxygen 93 Percent using a suitable paramagnetic analyzer. Acceptance criteria: 90.0%–96.0% oxygen by volume
4 IMPURITIES
Change to read:
See Impurities Testing in Medical Gases (ERR 1-Nov-2020) 〈413〉. The detector tubes called for in the following tests are listed in Reagents, Indicators, and Solutions.
4.1 Carbon Dioxide
Sample: Detector tube manufacturer's recommended volume ±5% of Oxygen 93 Percent
Analysis: Pass the Sample through a carbon dioxide detector tube at the rate specified for the tube by the detector tube manufacturer. Acceptance criteria: NMT 300 ppm
4.2 Carbon Monoxide
Sample: Detector tube manufacturer's recommended volume ±5% of Oxygen 93 Percent
Analysis: Pass the Sample through a carbon monoxide detector tube at the rate specified for the tube by the detector tube manufacturer. Acceptance criteria: NMT 10 ppm
5 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in pressurized containers. Container connections shall be appropriate for Oxygen 93 Percent. Adaptors shall not be used to connect containers to patient use supply system piping or equipment.
Labeling: If Oxygen 93 Percent is piped from a remote location to the patient point of use, label each outlet “Oxygen 93 Percent”.