Oxygen
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
O2 32.00
Oxygen CAS RN®: 7782-44-7; UNII: S88TT14065.
1 DEFINITION
Oxygen contains NLT 99.0% of oxygen (O2) by volume.
2 IDENTIFICATION
A. The paramagnetic signal exhibited by the Sample gas in the Assay confirms the presence of oxygen.
B. The Sample gas in the Assay meets the Acceptance criteria.
3 ASSAY
Procedure
The certified standards called for in the following test are listed in Reagents, Indicators, and Solutions.
Zero gas: Nitrogen certified standard
Span gas: Oxygen certified standard
Sample gas: Oxygen under test
Mode: Paramagnetic oxygen measurement (see Medical Gases Assay 〈415〉)
Analysis: Determine the concentration of oxygen in percentage by volume in Oxygen using a suitable paramagnetic analyzer. Acceptance criteria: NLT 99.0% by volume
4 IMPURITIES
Change to read:
See Impurities Testing in Medical Gases (ERR 1-Nov-2020) 〈413〉. The detector tubes called for in the following tests are listed in Reagents, Indicators, and Solutions. If the label indicates that Oxygen is produced by the air-liquefaction process, then the Impurities tests are not required.
4.1 Limit of Carbon Dioxide
Sample: Detector tube manufacturer's recommended volume, ±5% of Oxygen
Analysis: Pass the Sample through a carbon dioxide detector tube at the rate specified for the tube by the detector tube manufacturer. Acceptance criteria: NMT 300 ppm
4.2 Limit of Carbon Monoxide
Sample: Detector tube manufacturer's recommended volume, ±5% of Oxygen
Analysis: Pass the Sample through a carbon monoxide detector tube at the rate specified for the tube by the detector tube manufacturer. Acceptance criteria: NMT 10 ppm
5 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in pressurized containers. Container connections shall be appropriate for Oxygen. Adaptors shall not be used to connect containers to patient-use supply system piping or equipment.
Labeling: Label states if Oxygen was produced by the air-liquefaction process. Where it is piped directly from the cylinder or storage tank to the patient point of use, label each outlet “Oxygen”.

