Oxycodone Hydrochloride Tablets
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Oxycodone hydrochloride Tablets contain NLT 90.0% and NMT 110.0% of the labeled amount of oxycodone hydrochloride (C18H21NO4 · HCl).
2 IDENTIFICATION
A. The retention time of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay. Add the following:
B. The UV spectrum of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay. (USP 1- Dec-2023)
3 ASSAY
Change to read:
Procedure
Solution A: Dissolve 3.48 g of dibasic potassium phosphate in 1 L of water. Adjust with phosphoric acid to a pH of 6.7.
Solution B: Acetonitrile
Mobile phase: See Table 1.
Table 1
Time (min) | Solution A (%) | Solution B (%) |
0 | 90 | 10 |
1 | 90 | 10 |
12 | 50 | 50 |
12.1 | 90 | 10 |
15 | 90 | 10 |
Diluent: 0.1% (v/v) phosphoric acid in water
Standard solution: 0.1 mg/mL of USP Oxycodone Hydrochloride RS in Diluent
Sample solution: Nominally 0.1 mg/mL of oxycodone hydrochloride in Diluent, prepared as follows. Transfer a quantity of finely powdered Tablets (NLT 20), nominally equivalent to 5 mg of oxycodone hydrochloride, to a 50-mL volumetric flask. Add 40 mL of Diluent, sonicate for about 5 min. Cool to room temperature, and agitate on a mechanical shaker for about 15 min. Dilute with Diluent to volume. Centrifuge and use the clear supernatant.
3.1 Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 206 nm. For Identification B, use a diode array detector in the range of 200–400 nm.
Column: 4.6-mm × 10-cm; 3-µm packing L1
Column temperature: 45°
Flow rate: 0.7 mL/min
Injection volume: 5 µL
3.2 System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: NMT 2.0
Relative standard deviation: NMT 1.0%
3.3 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of oxycodone hydrochloride (C18H21NO4 · HCl) in the portion of Tablets taken:
Result = (rU /rS) × (CS /CU) × 100
rU = peak response of oxycodone from the Sample solution
rS = peak response of oxycodone from the Standard solution
CS = concentration of USP Oxycodone Hydrochloride RS in the Standard solution (mg/mL)
CU = nominal concentration of oxycodone hydrochloride in the Sample solution (mg/mL)
(USP 1-Dec-2023)
Acceptance criteria: 90.0%–110.0%
4 PERFORMANCE TESTS
Change to read:
Dissolution 〈711〉
Medium: Water; 500 mL
Apparatus 2: 50 rpm
Time: 45min
Standard solution: A known concentration of USP Oxycodone Hydrochloride RS in water (USP 1-Dec-2023) Sample solution: A portion of the solution under test. Dilute with Medium as needed.
4.1 Instrumental conditions
(See Ultraviolet-Visible Spectroscopy 〈857〉.)
Mode: UV
Analytical wavelength: Maximum at about 225 nm
4.2 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of oxycodone hydrochloride (C18H21NO4 · HCl) dissolved:
Result = (AU /AS ) × (CS /L) × V × D × 100
AU = absorbance of the Sample solution
AS = absorbance of the Standard solution
CS = concentration of USP Oxycodone Hydrochloride RS in the Standard solution (mg/mL)
L = label claim (mg/Tablet)
V = volume of Medium, 500 mL
D = dilution factor of the Sample solution, if applicable
(USP 1-Dec-2023)
Tolerances: NLT 70% (Q) of the labeled amount of oxycodone hydrochloride (C18H21NO4 · HCl) is dissolved.
Uniformity of Dosage Units 〈905〉: Meet the requirements
Add the following:
5 IMPURITIES
Organic Impurities
Solution A, Solution B, Mobile phase, Diluent, and Chromatographic system: Proceed as directed in the Assay.
System suitability solution: 1.0 mg/mL of USP Oxycodone Hydrochloride RS, 0.01 mg/mL each of USP Hydrocodone RS and USP Oxycodone Related Compound A RS in Diluent
Standard solution: 2 μg/mL each of USP Oxycodone Hydrochloride RS and USP Oxycodone Related Compound B RS in Diluent Sensitivity solution: 1 μg/mL each of USP Oxycodone Hydrochloride RS and USP Oxycodone Related Compound B RS in Diluent from Standard solution
Sample solution: Nominally 1.0 mg/mL of oxycodone hydrochloride in Diluent, prepared as follows. Transfer a quantity of finely powdered Tablets (NLT 20), nominally equivalent to 20 mg of oxycodone hydrochloride, to a 20-mL volumetric flask. Add 15 mL of Diluent, and sonicate for about 5 min. Cool to room temperature, and agitate on a mechanical shaker for about 15 min. Dilute with Diluent to volume. Centrifuge and use the clear supernatant.
System suitability
Samples: System suitability solution, Standard solution, and Sensitivity solution
[Note—The relative retention times for hydrocodone and oxycodone related compound A are about 1.2 and 1.3, respectively.]
5.1 Suitability requirements
Resolution: NLT 1.5 between hydrocodone and oxycodone related compound A, System suitability solution
Relative standard deviation: NMT 5.0% for oxycodone and oxycodone related compound B, Standard solution
Signal-to-noise ratio: NLT 10 for oxycodone and oxycodone related compound B, Sensitivity solution
5.2 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of oxycodone related compound B in the portion of Tablets taken:
Result = (rU /rS) × (CS /CU) × 100
rU = peak response of oxycodone related compound B from the Sample solution
rS = peak response of oxycodone related compound B from the Standard solution
CS = concentration of USP Oxycodone Related Compound B RS in the Standard solution (μg/mL)
CU = nominal concentration of oxycodone hydrochloride in the Sample solution (μg/mL)
Calculate the percentage of any unspecified degradation product in the portion of Tablets taken:
Result = (rU /rS ) × (CS /CU ) × 100
rU = peak response of any unspecified degradation product from the Sample solution
rS = peak response of oxycodone from the Standard solution
CS = concentration of USP Oxycodone Hydrochloride RS in the Standard solution (μg/mL)
CU = nominal concentration of oxycodone hydrochloride in the Sample solution (μg/mL)
Acceptance criteria: See Table 2. The reporting threshold is 0.1%.
Table 2
Name | Relative Retention Time | Acceptance Criteria, NMT (%) |
Oxycodone related compound B | 0.8 | 0.2 |
Oxycodone | 1.0 | — |
Any unspecified degradation product | — | 0.2 |
Total degradation products | — | 0.5 |
(USP 1-Dec-2023)
6 ADDITIONAL REQUIREMENTS
Change to read:
Packaging and Storage: Preserve in tight, light-resistant containers. Store at controlled room temperature. (USP 1-Dec-2023) Change to read:
USP Reference Standards 〈11〉
USP Hydrocodone RS
4,5α-Epoxy-3-methoxy-17-methylmorphinan-6-one.
C18H21NO3 299.37
USP Oxycodone Hydrochloride RS
USP Oxycodone Related Compound A RS
4,5α-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-7-ene-6-one.
C18H19NO4 313.35
USP Oxycodone Related Compound B RS
4,5α-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-6-one N-oxide.
C18H21NO5 331.37 (USP 1-Dec-2023)

