Oxycodone Hydrochloride Oral Solution
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Oxycodone hydrochloride Oral Solution contains NLT 90.0% and NMT 110.0% of the labeled amount of oxycodone hydrochloride (C18H21NO4 · HCl).
2 IDENTIFICATION
Change to read:
A. The UV spectrum of the oxycodone peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.(USP 1-May-2021)
Delete the following:
B. Thin-Layer Chromatography
Standard solution: Evaporate 5 mL of the Standard solution obtained from Identification test A just to dryness. Dissolve the residue in 1.0 mL of chloroform.
Sample solution: Evaporate 5 mL of the Sample solution obtained from Identification test A just to dryness. Dissolve the residue in 1.0 mL of chloroform.
2.1 Chromatographic system
(See Chromatography 〈621〉, Thin-Layer Chromatography.)
Mode: TLC
Adsorbent: 0.25-mm layer of chromatographic silica gel mixture
Application volume: 20 µL
Developing solvent system: Acetone, toluene, ether, and ammonium hydroxide (6:4:1:0.3)
2.2 Analysis
Samples: Standard solution and Sample solution
Develop the plate until the solvent front has moved about three-fourths of the length of the plate, remove it, mark the solvent front, allow the solvent to evaporate, and spray with iodoplatinate TS.
Acceptance criteria: The principal spot from the Sample solution corresponds in color, size, and R value to that from the solution from the F
3 Standard solution, and no other spots are observed. (USP 1-May-2021)
Change to read:
B. (USP 1-May-2021) The retention time of the oxycodone peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
ASSAY
Change to read:
Procedure
Mobile phase: Acetonitrile, 0.01 M sodium 1-hexanesulfonate, and glacial acetic acid (25:74:1). Adjust with 5 N sodium hydroxide to a pH of 3.5.
Standard solution: 0.045 mg/mL of USP Oxycodone RS in Mobile phase
Sample solution: Nominally 0.05 mg/mL of oxycodone hydrochloride in Mobile phase from Oral Solution. Pass a portion of this mixture through a filter of 0.5-µm or finer pore size, and use the clear filtrate as the Sample solution.
3.1 Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 280 nm.For Identification A, use a diode array detector in the range of 200–400 nm. (USP 1-May-2021) Column: 4.6-mm × 15-cm; 5-µm packing L1
Flow rate: 1.7 mL/min
Injection volume: 10 µL
3.2 System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: NMT 2.0
Relative standard deviation: NMT 2.0%
3.3 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of oxycodone hydrochloride (C18H21NO4 · HCl) in the portion of Oral Solution taken:
Result = (rU /rS) × (CS /CU) × (Mr1 /Mr2) × 100
rU = peak response from the Sample solution
rS = peak response from the Standard solution
CS = concentration of USP Oxycodone RS in the Standard solution (mg/mL)
CU = nominal concentration of oxycodone hydrochloride in the Sample solution (mg/mL)
Mr1 = molecular weight of oxycodone hydrochloride, 351.82
Mr2 = molecular weight of oxycodone free base, 315.37
Acceptance criteria: 90.0%–110.0%
4 IMPURITIES
Add the following:
Organic Impurities
Buffer: Dissolve 0.54 g of monobasic potassium phosphate in 1000 mL of water. Adjust with hydrochloric acid to a pH of 3.5. Add 6.5 g of anhydrous octanesulfonic acid sodium salt and readjust with hydrochloric acid to a pH of 3.5.
Solution A: Acetonitrile and Buffer (8:92)
Solution B: Acetonitrile and Buffer (45:55)
Mobile phase: See Table 1.
Table 1
Time (min) | Solution A (%) | Solution B (%) |
0 | 100 | 0 |
8 | 78 | 22 |
18 | 78 | 22 |
47 | 0 | 100 |
47.01 | 100 | 0 |
55 | 100 | 0 |
Diluent: Acetonitrile and 0.1 N hydrochloric acid (20:80)
System suitability solution: 0.45 mg/mL of USP Oxycodone RS and 0.001 mg/mL of USP Oxycodone Related Compound A RS in Diluent Sensitivity solution: 0.225 µg/mL of USP Oxycodone RS in Diluent
Standard solution: 0.0045 mg/mL of USP Oxycodone RS in Diluent
Sample solution: Nominally 0.5 mg/mL of oxycodone hydrochloride from Oral Solution in Diluent
4.1 Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 214 nm
Column: 4.6-mm × 10-cm; 2.6-µm packing L1
Column temperature: 45°
Flow rate: 1.0 mL/min
Injection volume: 10 µL
4.2 System suitability
Samples: System suitability solution, Sensitivity solution, and Standard solution
[Note—The relative retention times for oxycodone related compound A and 7-methyloxycodone (4,5α-Epoxy-14-hydroxy-3-methoxy-7,17- dimethylmorphinan-6-one) are 1.05 and 1.17, respectively.]
Suitability requirements
Resolution: NLT 2.0 between oxycodone and oxycodone related compound A, System suitability solution
Relative standard deviation: NMT 5.0%, Standard solution
Signal-to-noise ratio: NLT 10, Sensitivity solution
4.3 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of each individual degradation product in the portion of Oral Solution taken:
Result = (rU /rS) × (CS /CU) × (Mr1 /Mr2) × (1/F) × 100
U S S U r1 r2
rU = peak response of each individual degradation product from the Sample solution
rS = peak response of oxycodone from the Standard solution
CS = concentration of USP Oxycodone RS in Standard solution (mg/mL)
CU = nominal concentration of oxycodone hydrochloride in the Sample solution (mg/mL)
Mr1 = molecular weight of oxycodone hydrochloride, 351.82
Mr2 = molecular weight of oxycodone free base, 315.37
F = relative response factor (see Table 2)
Acceptance criteria: See Table 2.
Table 2
Name | Relative Retention Time | Relative Response Factor | Acceptance Criteria, NMT (%) |
Oxycodone N-oxide (oxycodone related compound B) | 0.42 | 1.14 | 1.0 |
Oxycodone | 1.00 | — | — |
Any unspecified degradation product | — | 1.00 | 0.2 |
Total degradation products | — | — | 2.0 |
(USP 1-May-2021)
5 OTHER COMPONENTS
Alcohol Determination 〈611〉, Procedures, Method II (if present): 85.0%–115.0% of the labeled amount of alcohol (C H OH), determined by the 2 5
gas–liquid chromatographic method, using acetone as the internal standard
6 PERFORMANCE TESTS
Uniformity of Dosage Units 〈905〉
For Oral Solution packaged in single-unit containers: Meets the requirements
Deliverable Volume 〈698〉
For Oral Solution packaged in multiple-unit containers: Meets the requirements
7 SPECIFIC TESTS
pH 〈791〉: 1.4–4.6
ADDITIONAL REQUIREMENTS
Change to read:
Packaging and Storage: Preserve in tight, light-resistant containers. Store at controlled room temperature. (USP 1-May-2021)
Change to read:
USP Reference Standards 〈11〉
USP Oxycodone RS
USP Oxycodone Related Compound A RS
4,5α-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-7-ene-6-one.
C18H19NO4 313.35 (USP 1-May-2021)

