Oxycodone Hydrochloride
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
C18H21NO4 · HCl 351.82
Morphinan-6-one, 4,5-epoxy-14-hydroxy-3-methoxy-17- methyl-, hydrochloride, (5α)-;
4,5α-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-6- one hydrochloride CAS RN®: 124-90-3; UNII: C1ENJ2TE6C.
1 DEFINITION
Oxycodone hydrochloride contains NLT 97.0% and NMT 103.0% of oxycodone hydrochloride (C18H21NO4 · HCl), calculated on the anhydrous, solvent-free basis.
2 IDENTIFICATION
Change to read:
A. Spectroscopic Identification Tests 〈197〉, Infrared Spectroscopy: 197A or 197K (CN 1-May-2020)
Sample: Dissolve 250 mg of Oxycodone Hydrochloride in 25 mL of water. Render 25 mL of the resulting solution with 6 N ammonium hydroxide. Allow the mixture to stand until a precipitate is formed. Filter, wash the precipitate with 50 mL of cold water, and dry at 105° for 2 h.
Acceptance criteria: Meets the requirements
B. The retention time of the oxycodone peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
3 ASSAY
Procedure
Mobile phase: 0.005 M sodium 1-hexanesulfonate, methanol, triethylamine, and phosphoric acid (900:100:2:5). Adjust with 50% sodium hydroxide solution to a pH of 2.5 ± 0.1 and filter.
System suitability solution: 13 µg/mL of USP Codeine Phosphate RS and 9 µg/mL of USP Oxycodone RS in Mobile phase Standard solution: 0.9 mg/mL of USP Oxycodone RS in Mobile phase
Sample solution: 1 mg/mL of Oxycodone Hydrochloride in Mobile phase. [Note—Pass a portion of this solution through a filter of 0.5-µm or finer pore size, and use the filtrate as the Sample solution.]
3.1 Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 206 nm
Column: 3.9-mm × 15-cm; 4-µm packing L7
Column temperature: 50°
Flow rate: 1.5 mL/min
Injection volume: 10 µL
Run time: NLT 2 times the retention time of oxycodone
3.2 System suitability
Samples: System suitability solution and Standard solution
[Note—The relative retention times for codeine and oxycodone are about 0.8 and 1.0, respectively.]
Suitability requirements
Resolution: NLT 3.0 between codeine and oxycodone, System suitability solution
Tailing factor: 0.75–1.25, Standard solution
Relative standard deviation: NMT 1.10%, Standard solution
3.3 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of oxycodone hydrochloride (C18H21NO4 · HCl) in the portion of Oxycodone Hydrochloride taken:
Result = (rU /rS) × (CS /CU) × (Mr1 /Mr2) × 100
rU = peak response from the Sample solution
rS = peak response from the Standard solution
CS = concentration of USP Oxycodone RS in the Standard solution (mg/mL)
CU = concentration of Oxycodone Hydrochloride in the Sample solution (mg/mL)
Mr1 = molecular weight of oxycodone hydrochloride, 351.82
Mr2 = molecular weight of oxycodone base, 315.37
Acceptance criteria: 97.0%–103.0% on the anhydrous, solvent-free basis
4 IMPURITIES
Residue on Ignition 〈281〉: NMT 0.05%
[Note—Use of sulfuric acid is omitted.]
[Note—On the basis of the synthetic route, perform either Procedure 1 or Procedure 2. Procedure 1 is recommended if 8β-hydroxyoxycodone (7,8-dihydro-8β-14-dihydroxycodeinone) is a potential impurity.]
4.1 Organic Impurities, Procedure 1
Mobile phase, System suitability solution, Standard solution, Sample solution, Chromatographic system, and System suitability: Proceed as directed in the Assay.
Analysis
Sample: Sample solution
Calculate the percentage of each impurity in the portion of Oxycodone Hydrochloride taken:
Result = (rU /rT ) × 100
rU = peak response of each impurity from the Sample solution
rT = sum of the responses of all the peaks from the Sample solution
Acceptance criteria: See Table 1.
Table 1
Name | Relative Retention Time | Acceptance Criteria, NMT (%) |
Oxymorphone | 0.31 | 0.15 |
Noroxymorphonea | 0.33 | 0.15 |
10-Hydroxyoxycodoneb | 0.53 | 0.15 |
6-α Oxycodolc | 0.67 | 0.25 |
8β-Hydroxyoxycodone (7,8-dihydro-8β-14-dihydroxycodeinone)d | 0.71 | 0.15 |
Oxycodone | 1.00 | — |
Hydrocodone | 1.19 | 0.15 |
Individual unspecified impurity | — | 0.10 |
Total impurities | — | 2.0 |
a4,5α-Epoxy-3,14-dihydroxymorphinan-6-one.
b 4,5α-Epoxy-10α,14-dihydroxy-3-methoxy-17-methylmorphinan-6-one.
c 4,5α-Epoxy-3-methoxy-17-methylmorphinan-6α,14-diol.
d 4,5α-Epoxy-8β,14-dihydroxy-3-methoxy-17-methylmorphinan-6-one.
4.2 Organic Impurities, Procedure 2
Buffer: Mix 4.0 mL of heptafluorobutyric acid with 2000 mL of water and adjust with ammonium hydroxide to a pH of 2.3 ± 0.1.
Solution A: Methanol and Buffer (23:77)
Solution B: Methanol, tetrahydrofuran, and Buffer (20:3:77)
Mobile phase: See Table 2.
Table 2
Time (min) | Solution A (%) | Solution B (%) |
0 | 100 | 0 |
2 | 100 | 0 |
30 | 0 | 100 |
55 | 0 | 100 |
55.1 | 100 | 0 |
65 | 100 | 0 |
Diluent: Trifluoroacetic acid and water (3:1000)
System suitability solution: 0.0067 mg/mL each of USP Hydrocodone RS and USP Oxycodone Related Compound A RS, and 3.0 mg/mL of USP Oxycodone Hydrochloride RS in Diluent
Standard solution: 0.0067 mg/mL of USP Hydrocodone RS in Diluent
Sample solution: 3.0 mg/mL of Oxycodone Hydrochloride in Diluent
4.2.1 Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 280 nm
Column: 4.6-mm × 25-cm; 3-µm packing L1
Column temperature: 38°
Flow rate: 0.8 mL/min
Injection volume: 50 µL
4.2.2 System suitability
Samples: System suitability solution and Standard solution
Suitability requirements
Resolution: NLT 2.0 between oxycodone and hydrocodone; NLT 1.0 between hydrocodone and oxycodone related compound A, System suitability solution
Relative standard deviation: NMT 5.0%, Standard solution
4.2.3 Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of each impurity in the portion of Oxycodone Hydrochloride taken:
Result = (rU /rS) × (CS /CU) × (Mr1 /Mr2) × (1/F) × 100
rU = peak response of each impurity from the Sample solution
rS = peak response of hydrocodone from the Standard solution
CS = concentration of USP Hydrocodone RS in the Standard solution (mg/mL)
CU = concentration of Oxycodone Hydrochloride in the Sample solution (mg/mL)
Mr1 = molecular weight of hydrocodone hydrochloride, 335.83
Mr2 = molecular weight of hydrocodone, 299.36
F = relative response factor (see Table 3)
Acceptance criteria: See Table 3. Disregard any peaks below 0.03%.
Table 3
Name | Relative Retention Time | Relative Response Factor | Acceptance Criteria, NMT (%) |
Oxymorphone hydrochloride | 0.54 | 0.93 | 0.15 |
1-Hydroxyoxycodone hydrochloridea | 0.69 | 1.00 | 0.15 |
6-Oxycodol hydrochlorideb | 0.79 | 1.16 | 0.25 |
Oxycodone hydrochloride | 1.00 | — | — |
Hydrocodone hydrochloride | 1.14 | 1.00 | 0.50 |
14-Hydroxycodeinone hydrochloride (oxycodone related compound A hydrochloride)c | 1.18 | 0.99 | 0.25 |
Noroxycodone hydrochlorided | 1.26 | 0.94 | 0.50 |
Individual unspecified impurity | — | — | 0.10 |
Total impurities | — | — | 1.5 |
a 4,5α-Epoxy-1,14-dihydroxy-3-methoxy-17-methylmorphinan-6-one hydrochloride.
b 4,5α-Epoxy-3-methoxy-17-methylmorphinan-6,14-diol hydrochloride.
c 4,5α-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-7-ene-6-one hydrochloride (oxycodone related compound A hydrochloride salt).
d 4,5α-Epoxy-14-hydroxy-3-methoxymorphinan-6-one hydrochloride.
5 SPECIFIC TESTS
Content of Chloride
Sample solution: 6 mg/mL of Oxycodone Hydrochloride in methanol
Analysis: To 50 mL of the Sample solution, add 5 mL of glacial acetic acid and titrate with 0.1 N silver nitrate VS, determining the endpoint potentiometrically. Each milliliter of 0.1 N silver nitrate is equivalent to 3.545 mg of chloride.
Acceptance criteria: 9.8%–10.4% on the anhydrous, solvent-free basis
Optical Rotation 〈781S〉, Procedures, Specific Rotation
Sample solution: 25 mg/mL of Oxycodone Hydrochloride in water on the anhydrous, solvent-free basis
Acceptance criteria: −137° to −149°
Water Determination 〈921〉, Method I: NMT 7.0%
6 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in tight containers.
Labeling: The label states with which Organic Impurities procedure the article complies if Organic Impurities, Procedure 1 is not used.
USP Reference Standards 〈11〉
USP Codeine Phosphate RS
USP Hydrocodone RS
USP Oxycodone RS
USP Oxycodone Hydrochloride RS
USP Oxycodone Related Compound A RS
Also known as 14-Hydroxycodeinone;
4,5α-Epoxy-14-hydroxy-3-methoxy-17-methylmorphinan-7-ene-6-one.
C18H19NO4 313.35

