Oxaliplatin for Injection
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Oxaliplatin for Injection is a sterile, lyophilized mixture of Oxaliplatin and Lactose Monohydrate. It contains NLT 90.0% and NMT 110.0% of the labeled amount of oxaliplatin (C8H14N2O4Pt).
2 IDENTIFICATION
A. The retention time of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
Add the following:
B. The UV spectrum of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay. 2S (USP41)
3 ASSAY
Change to read:
Procedure
[Note - Use vigorous shaking and very brief sonication to dissolve the substance to be examined. Inject the Sample solution within 20 min of preparation. Use polypropylene HPLC autosampler vials.]
Acidified water: Adjust water with phosphoric acid to a pH of 3.0.
Mobile phase: Acetonitrile and Acidified water (1:99)
System suitability solution: 0.1 mg/mL each of USP Oxaliplatin RS and USP Oxaliplatin System Suitability RS in water. [Note - USP Oxaliplatin System Suitability RS is [SP-4-2-(1R-trans)]-(1,2-cyclohexanediamine-N,N′) dichloridoplatinum(II).]
Standard solution: 0.1 mg/mL of USP Oxaliplatin RS in water
Sample solution: Nominally equivalent to 0.1 mg/mL of oxaliplatin obtained by constituting a suitable number of vials of Oxaliplatin for Injection with the appropriate amount of water
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 210 nm. For Identification B, use a diode array detector in the range of 200–400 nm. 2S (USP41)
Column: 4.6-mm × 25-cm; 5-μm packing L1
Column temperature: 40°
Flow rate: 1.2 mL/min
Injection volume: 20 μL
System suitability
Samples: System suitability solution and Standard solution
[Note - The relative retention times for [SP-4-2-(1R-trans)]-(1,2-cyclohexanediamine-N,N′) dichloridoplatinum(II) and oxaliplatin are about 0.9 and 1.0, respectively.]
Suitability requirements
Resolution: NLT 2.0 between the peaks of [SP-4-2-(1R-trans)]-(1,2-cyclohexanediamine-N,N′) dichloridoplatinum(II) and oxaliplatin, System suitability solution
Tailing factor: NMT 2.0 for the oxaliplatin peak, System suitability solution
Relative standard deviation: NMT 1.0%, Standard solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of oxaliplatin (C8H14N2O4Pt) in the portion of Oxaliplatin for Injection taken:
Result = (rU/rS) × (CS/CU) × 100
rU = peak response from the Sample solution
rS = peak response from the Standard solution
CS = concentration of USP Oxaliplatin RS in the Standard solution (mg/mL)
CU = nominal concentration of oxaliplatin in the Sample solution (mg/mL)
Acceptance criteria: 90.0%–110.0%
4 PERFORMANCE TESTS
Uniformity of Dosage Units 〈905〉: Meets the requirements
5 IMPURITIES
Change to read:
5.1 Limit of Oxalic Acid
[Note - Use vigorous shaking and very brief sonication to dissolve the substance to be examined. Inject the Sample solution within 20 min of preparation. Use polypropylene HPLC autosampler vials.]
Buffer: Add 1.36 g of monobasic potassium phosphate 2S (USP41) to 10 mL of 10% tetrabutylammonium hydroxide in water, and dilute with water to 1000 mL. Adjust with phosphoric acid to a pH of 6.0.
Mobile phase: Acetonitrile and Buffer (1:4)
Standard stock solution: 0.06 mg/mL of USP Oxaliplatin Related Compound A RS in water. [Note - USP Oxaliplatin Related Compound A RS is available as oxalic acid dihydrate.]
Standard solution: 15 μg/mL of USP Oxaliplatin Related Compound A RS in water, from the Standard stock solution
System suitability stock solution: 0.05 mg/mL of sodium nitrate in water
System suitability solution: 1.0 μg/mL of sodium nitrate and 15 μg/mL of oxaliplatin related compound A in water, from the System suitability stock solution and Standard stock solution, respectively
Sensitivity solution: 1.5 μg/mL of USP Oxaliplatin Related Compound A RS in water, from the Standard solution 2S (USP41)
Sample solution: Nominally equivalent to 2.0 mg/mL of oxaliplatin in water from Oxaliplatin for Injection
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 205 nm
Column: 4.6-mm × 25-cm; 5-μm packing L1
Column temperature: 40°
Flow rate: 2 mL/min
Injection volume: 20 μL
System suitability
Samples: Standard solution, System suitability solution, and Sensitivity solution
[Note - The relative retention times for the sodium nitrate and oxalic acid peaks are about 0.6 and 1.0, respectively.]
Suitability requirements
Resolution: NLT 2.0 between the oxalic acid and sodium nitrate peaks, System suitability solution
Relative standard deviation: NMT 3.0% for the oxalic acid peak, Standard solution
Signal-to-noise ratio: NLT 10, Sensitivity solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of oxalic acid in the portion of Oxaliplatin for Injection taken:
Result = (rU/rS) × (CS/CU) × (Mr1/Mr2) × 100
rU = peak response of oxalic acid from the Sample solution
rS = peak response of oxalic acid from the Standard solution
CS = concentration of USP Oxaliplatin Related Compound A RS in the Standard solution (mg/mL)
CU = nominal concentration of oxaliplatin in the Sample solution (mg/mL)
Mr1 = molecular weight of anhydrous oxalic acid, 90.03
Mr2 = molecular weight of USP Oxaliplatin Related Compound A RS, 126.07
Acceptance criteria: NMT 0.5%
Change to read:
5.2 Limit of (SP-4-2)-Diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum
[Note - Use vigorous shaking and very brief sonication to dissolve the substance to be examined. Inject the Sample solution within 20 min of preparation. Use polypropylene HPLC autosampler vials.]
Buffer: Dissolve 1.36 g of monobasic potassium phosphate 2S (USP41) and 1 g of sodium 1-heptanesufonate in 1 L of water. Adjust with phosphoric acid to a pH of 3.0.
Mobile phase: Acetonitrile and Buffer (1:4)
System suitability solution: 2 mg/mL of USP Oxaliplatin RS in 0.005 M sodium hydroxide. Allow this solution to stand at room temperature for at least 5 days. [Note - Sonicate if necessary.] Transfer 5 mL of this solution into a 50-mL volumetric flask, and dilute with water to volume. [Note - The preparation of the System suitability solution forms diaquodiaminocyclohexaneplatinum dimer and (SP-4-2) diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum.]
Standard solution: 0.0125 mg/mL of USP Oxaliplatin Related Compound B RS prepared as follows. Transfer USP Oxaliplatin Related Compound B RS to a suitable volumetric flask, add 25% of the final volume of methanol, and sonicate for approximately 30 min to dissolve.
Allow to cool, if necessary, and dilute with water to volume. [Note - When preparing the solution, USP Oxaliplatin Related Compound B RS is converted to (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′] platinum.]
Sample solution: Use the Sample solution from the test for Limit of Oxalic Acid.
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 215 nm
Column: 4.6-mm × 25-cm; 5-μm packing L1
Column temperature: 40°
Flow rate: 2 mL/min
Injection volume: 20 μL
System suitability
Samples: System suitability solution and Standard solution
[Note - The relative retention times for (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum and diaquodiaminocyclohexaneplatinum dimer are about 1.0 and 1.5, respectively.]
Suitability requirements
Resolution: NLT 2.0 between (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum and diaquodiaminocyclohexaneplatinum dimer, System suitability solution
Relative standard deviation: NMT 3.0%, Standard solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum in the portion of Oxaliplatin for Injection
taken:
Result = (rU/rS) × (CS/CU) × (Mr1/Mr2) × 100
rU = peak response of (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum from the Sample solution
rS = peak response of (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum from the Standard solution
CS = concentration of USP Oxaliplatin Related Compound B RS in the Standard solution (mg/mL)
CU = nominal concentration of oxaliplatin in the Sample solution (mg/mL)
Mr1 = molecular weight of (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum, 345.30
Mr2 = molecular weight of USP Oxaliplatin Related Compound B RS, 433.28
Acceptance criteria: NMT 0.5%
Change to read:
5.3 Limit of Related Compound C and Unspecified Impurities
[Note-Use vigorous shaking and very brief sonication to dissolve the substance to be examined. Inject the Sample solution within 20 min of preparation. Use polypropylene HPLC autosampler vials.]
Mobile phase: Prepare as directed in the Assay.
Standard stock solution: 0.1 mg/mL each of USP Oxaliplatin RS and USP Oxaliplatin Related Compound C RS in water
Standard solution: 0.01 mg/mL each of USP Oxaliplatin RS and USP Oxaliplatin Related Compound C RS in water, from the Standard stock solution
System suitability stock solution: 0.1 mg/mL of USP Oxaliplatin System Suitability RS in methanol. Sonicate for approximately 10 min to aid the dissolution.
System suitability solution: Transfer 10 mL each of the Standard stock solution and the System suitability stock solution into a 100-mL volumetric flask, and dilute with water to volume.
Sample solution: Use the Sample solution from the test for Limit of Oxalic Acid.
Chromatographic system: Proceed as directed in the Assay, except for the Injection volume.
Injection volume: 10 μL
System suitability
Samples: Standard solution and System suitability solution
Suitability requirements
Resolution: NLT 2.0 between [SP-4-2-(1R-trans)]-(1,2-cyclohexanediamine-N,N′) dichloridoplatinum(II) and oxaliplatin, System suitability solution
Tailing factor: NMT 2.0 for the oxaliplatin peak, System suitability solution
Relative standard deviation: NMT 3.0% each for the oxaliplatin and oxaliplatin related compound C peaks, Standard solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of oxaliplatin related compound C in the portion of Oxaliplatin for Injection taken:
Result = (rU/rS) × (CS/CU) × 100
rU = peak response of oxaliplatin related compound C from the Sample solution
rS = peak response of oxaliplatin related compound C from the Standard solution
CS = concentration of USP Oxaliplatin Related Compound C RS in the Standard solution (mg/mL)
CU = nominal concentration of oxaliplatin in the Sample solution (mg/mL)
Calculate the percentage of each unspecified impurity in the portion of Oxaliplatin for Injection taken:
Result = (rU/rS) × (CS/CU) × 100
rU = peak response of each impurity from the Sample solution
rS = peak response of oxaliplatin from the Standard solution
CS = concentration of USP Oxaliplatin RS in the Standard solution (mg/mL)
CU = nominal concentration of oxaliplatin in the Sample solution (mg/mL)
Acceptance criteria: See Table 1.
Table 1
| Name | Relative Retention Time | Acceptance Criteria, NMT (%) |
|---|---|---|
| Oxaliplatin related compound C2S (USP41) | 0.6 | 0.3 |
| [SP-4-2-(1R-trans)-(1,2-Cyclohexanediamine-N,N′) dichloridoplatinum(II)]ᵃ | 0.8 | – |
| Oxaliplatin | 1.0 | – |
| Any individual unspecified impurity | – | 0.2 |
| Total impuritiesᵇ | – | 1.5 |
a The relative retention time is included for system suitability purposes only.
bIncludes oxalic acid, (SP-4-2)-diaqua[(1R,2R)-cyclohexane-1,2-diamine-N,N′]platinum, oxaliplatin related compound C, and the total of the individual unspecified impurities.
6 SPECIFIC TESTS
pH 〈791〉: 4.0–7.0 using a polymer combination electrode, determined in a solution constituted as directed in the labeling
Particulate Matter in Injections 〈788〉: It meets the requirements for small-volume injections.
Constituted Solution: At the time of use, it meets the requirements in Injections and Implanted Drug Products 〈1〉, Product Quality Tests Common to Parenteral Dosage Forms, Specific Tests, Completeness and Clarity of Solutions.
Bacterial Endotoxins Test 〈85〉: NMT 1.0 USP Endotoxin Unit/mg of oxaliplatin
Sterility Tests 〈71〉: Meets the requirements
Water Determination 〈921〉, Method I: NMT 4.0%
Other Requirements: It meets the requirements in Injections and Implanted Drug Products 〈1〉.
7 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve as described in Packaging and Storage Requirements 〈659〉, Injection Packaging, Packaging for Constitution. Store at controlled room temperature.
Labeling: Label it to indicate that it is to be diluted with a suitable parenteral vehicle before intravenous infusion.
Change to read:
USP Reference Standards 〈11〉 (CN 1-May-2018)
USP Oxaliplatin RS 2S (USP41)
USP Oxaliplatin Related Compound A RS
Oxalic acid dihydrate.
C2H2O4 · 2H O 126.07
USP Oxaliplatin Related Compound B RS
[SP-4-2-(1R-trans)]-(1,2-Cyclohexanediamine-N,N′) dinitratoplatinum(II).
C6H14N4O6Pt 433.28
USP Oxaliplatin Related Compound C RS
[1R-trans-(1,2-Cyclohexanediamine-N,N′)]-trans-dihydroxido-[oxalato(2-)-O,O′]platinum(IV).
C8H16N2O6Pt 431.30
USP Oxaliplatin System Suitability RS
[SP-4-2-(1R-trans)]-(1,2-Cyclohexanediamine-N,N′) dichloridoplatinum(II).
C6H14Cl2N2Pt 380.17

