Ondansetron Injection
If you find any inaccurate information, please let us know by providing your feedback here

This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Ondansetron Injection is a sterile solution of Ondansetron Hydrochloride in Water for Injection or of Ondansetron in Water for Injection prepared with the aid of Hydrochloric Acid. It may contain suitable buffers and/or tonicity adjusting agents. It contains an amount of Ondansetron Hydrochloride equivalent to NLT 95.0% and NMT 105.0% of the labeled amount of ondansetron (C₁₈H₁₉N₃O).
2 IDENTIFICATION
A. The retention time of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
Add the following:
▲B. The UV spectrum of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.▲ (USP 1-May-2024)
3 ASSAY
Change to read:
Procedure
▲Solution A: 2.4 g/L of monobasic sodium phosphate anhydrous and 0.6 g/L of sodium 1-heptanesulfonate in water. Adjust with 0.5 N sodium hydroxide to a pH of 5.4.
Solution B: Acetonitrile
Mobile phase: See Table 1.
| Time (min) | Solution A (%) | Solution B (%) |
|---|---|---|
| 0 | 95 | 5 |
| 5 | 95 | 5 |
| 8 | 80 | 20 |
| 25 | 35 | 65 |
| 27 | 30 | 70 |
| 33 | 30 | 70 |
| 35 | 95 | 5 |
| 40 | 95 | 5 |
Diluent: Acetonitrile and water (30:70). To each liter of this solution, add 1 mL of formic acid.
Standard solution: 0.1 mg/mL of USP Ondansetron Hydrochloride RS in Diluent
Sample solution: Nominally 0.1 mg/mL of ondansetron from Injection in Diluent
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 216 nm. For Identification B, use a diode array detector in the range of 200–400 nm.
Column: 4.6-mm × 25-cm; 5-µm packing L11
Temperatures
Autosampler: 15°
Column: 40°
Flow rate: 1 mL/min
Injection volume: 10 µL
System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: NMT 2.0
Relative standard deviation: NMT 1.0% ▲ (USP 1-May-2024)
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of ondansetron (C₁₈H₁₉N₃O) in the portion of Injection taken:
Result = (rᵤ / rₛ) × (Cₛ / Cᵤ) × (M₁ / M₂) × 100
rᵤ = peak response of ondansetron from the Sample solution
rₛ = peak response of ondansetron from the Standard solution
Cₛ = concentration ▲ (USP 1-May-2024) of USP Ondansetron Hydrochloride RS in the Standard solution (mg/mL)
Cᵤ = nominal concentration of ondansetron in the Sample solution (mg/mL)
M₁ = molecular weight of ondansetron, ▲293.37▲ (USP 1-May-2024)
M₂ = molecular weight of anhydrous ondansetron hydrochloride, 329.83
Acceptance criteria: 95.0%–105.0%
4 IMPURITIES
Delete the following:
▲Limit of Ondansetron Related Compound D▲ (USP 1-May-2024)
Change to read:
Organic Impurities
▲Solution A, Solution B, Mobile phase, Diluent, and Chromatographic system: Proceed as directed in the Assay.
System suitability solution: 1 mg/mL of USP Ondansetron Hydrochloride RS and 0.002 mg/mL of USP Ondansetron Related Compound G in Diluent
Standard solution: 0.002 mg/mL each of USP Ondansetron Hydrochloride RS and USP Ondansetron Related Compound D RS in Diluent
Sensitivity solution: 0.001 mg/mL each of USP Ondansetron Hydrochloride RS and USP Ondansetron Related Compound D RS from the Standard solution in Diluent
Sample solution: Nominally 1.0 mg/mL of ondansetron from Injection in Diluent
System suitability
Samples: System suitability solution, Standard solution, and Sensitivity solution
[Note—The relative retention times in Table 2 are provided as information that could aid in peak assignment.]
| Name | Relative Retention Time |
|---|---|
| Imidazoleᵃ | 0.28 |
| Ondansetron related compound Fᵇ | 0.33 |
| Ondansetron related compound Aᶜ | 0.94 |
| Ondansetron | 1.00 |
| Ondansetron related compound G | 1.04 |
| Ondansetron related compound Cᵈ | 1.10 |
| Ondansetron related compound D | 1.20 |
ᵃ 1H-Imidazole.
ᵇ 2-Methyl-1H-imidazole.
ᶜ 3-[(Dimethylamino)methyl]-9-methyl-1,2,3,9-tetrahydro-4H-carbazol-4-one hydrochloride.
ᵈ 9-Methyl-1,2,3,9-tetrahydro-4H-carbazol-4-one.
Suitability requirements
Resolution: NLT 2.0 between ondansetron and ondansetron related compound G, System suitability solution
Relative standard deviation: NMT 5.0% for ondansetron, Standard solution
Signal-to-noise ratio: NLT 10 for ondansetron, Sensitivity solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of ondansetron related compound D in the portion of Injection taken:
Result = (rᵤ / rₛ) × (Cₛ / Cᵤ) × 100
rᵤ = peak response of ondansetron related compound D from the Sample solution
rₛ = peak response of ondansetron related compound D from the Standard solution
Cₛ = concentration of USP Ondansetron Related Compound D RS in the Standard solution (mg/mL)
Cᵤ = nominal concentration of ondansetron in the Sample solution (mg/mL)
Calculate the percentage of any other specified or unspecified degradation product in the portion of Injection taken:
Result = (rᵤ / rₛ) × (Cₛ / Cᵤ) × (1/F) × (M₁ / M₂) × 100
rᵤ = peak response of any other specified or any unspecified degradation product from the Sample solution
rₛ = peak response of ondansetron from the Standard solution
Cₛ = concentration of USP Ondansetron Hydrochloride RS in the Standard solution (mg/mL)
Cᵤ = nominal concentration of ondansetron in the Sample solution (mg/mL)
F = relative response factor (see Table 3)
M₁ = molecular weight of ondansetron, 293.37
M₂ = molecular weight of anhydrous ondansetron hydrochloride, 329.83
Acceptance criteria: See Table 3. The reporting threshold is 0.1%.
| Name | Relative Response Factor | Acceptance Criteria, NMT (%) |
|---|---|---|
| Ondansetron related compound A | 1.0 | 0.2 |
| Ondansetron related compound C | 1.5 | 0.2 |
| Ondansetron related compound D | — | 0.12 |
| Any unspecified degradation product | 1.0 | 0.2 |
| Total degradation products | — | 0.5 |
5 SPECIFIC TESTS
Change to read:
Bacterial Endotoxins Test 〈85〉: ▲Meets the requirements▲ (USP 1-May-2024)
Add the following:
▲• Sterility Tests 〈71〉: Meets the requirements▲ (USP 1-May-2024)
pH 〈791〉: 3.3–4.0
Particulate Matter in Injections 〈788〉: Meets the requirements for small-volume injections
Other Requirements: It meets the requirements under Injections and Implanted Drug Products 〈1〉.
6 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in single-dose or in multiple-dose containers, preferably of Type I glass, at a temperature between 2° and 30°, protected from light.
Change to read:
USP Reference Standards 〈11〉
USP Ondansetron Hydrochloride RS
▲ (USP 1-May-2024)
USP Ondansetron Related Compound D RS
9-Methyl-3-methylene-1,2,3,9-tetrahydro-4H-carbazol-4-one.
▲C₁₄H₁₃NO 211.26
USP Ondansetron Related Compound G
3-[(1H-Imidazole-1-yl)methyl]-9-methyl-1,2,3,9-tetrahydro-4H-carbazol-4-one.
C₁₇H₁₇N₃O 279.34▲ (USP 1-May-2024)

