Octocrylene
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
C24H27NO2 361.48
2-Propenoic acid, 2-cyano-3,3-diphenyl-, 2-ethylhexyl ester;
2-Ethylhexyl 2-cyano-3,3-diphenylacrylate
CAS RN®: 6197-30-4; UNII: 5A68WGF6WM
1 DEFINITION
Octocrylene contains NLT 95.0% and NMT 105.0% of octocrylene (C24H27NO2).
2 IDENTIFICATION
Change to read:
A. Spectroscopic Identification Tests 〈197〉, Ultraviolet-Visible Spectroscopy: 197U
Sample solution: 25 µg/mL in methanol
Analytical wavelength: 303 nm
Acceptance criteria: Absorptivities, calculated on the as-is basis, do not differ by more than 3.0%. (ERR 1-Sep-2023)
B. The retention time of major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
3 ASSAY
3.1 Procedure
Standard solution: 21 mg/mL of USP Octocrylene RS in acetone
Sample solution: 21 mg/mL of Octocrylene in acetone
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: GC
Detector: Flame ionization
Column: 0.32-mm × 60-m; coated with a 0.25-µm film of phase G1
Temperatures
Injection port: 300°
Detector: 300°
Column: See Table 1.
Table 1
| Initial Temperature (°) | Temperature Ramp (°/min) | Final Temperature (°) | Hold Time at Final Temperature (min) |
| 80 | 4 | 280 | 10 |
Carrier gas: Helium
Flow rate: 6 mL/min
Injection volume: 1 µL
Injection type: Split ratio, 30:1
System suitability
Sample: Standard solution
Suitability requirements
Resolution: NLT 1.0 between octocrylene and any other adjacent peak
Relative standard deviation: NMT 2.0%
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of octocrylene (C24H27NO2) in the portion of Octocrylene taken:
Result = (rU/rS) × (CS/CU) × 100
rU = peak response from the Sample solution
rS = peak response from the Standard solution
CS = concentration of USP Octocrylene RS in the Standard solution (mg/mL)
CU = concentration of Octocrylene in the Sample solution (mg/mL)
Acceptance criteria: 95.0%–105.0%
4 IMPURITIES
4.1 Organic Impurities
Standard solution, Sample solution, Chromatographic system, and System suitability: Proceed as directed in the Assay.
Analysis
Sample: Sample solution
Calculate the percentage of each impurity in the portion of Octocrylene taken:
Result = (rU/rT) × 100
rU = peak response of each impurity from the Sample solution
rT = sum of all the peak responses from the Sample solution
Acceptance criteria
Individual impurities: NMT 0.5%
Total impurities: NMT 2.0%
5 SPECIFIC TESTS
Specific Gravity 〈841〉: 1.045–1.055
Refractive Index 〈831〉: 1.561–1.571 at 20°
Acidity
Sample: 6 g of Octocrylene
Titrimetric system
(See Titrimetry 〈541〉.)
Mode: Direct titration
Titrant: 0.1 N sodium hydroxide
Endpoint detection: Visual
Analysis: To 60 mL of alcohol add 1 mL of phenolphthalein TS and sufficient Titrant to obtain a persistent pink color. Transfer 60 mL of this solution to a suitable container, add the Sample, mix, and titrate with Titrant until a persistent pink color is obtained.
Acceptance criteria: NMT 0.18 mL of Titrant/g is required
6 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in tight containers.
USP Reference Standards 〈11〉
USP Octocrylene RS

