Nitrogen
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
N₂ 28.01
Nitrogen CAS RN®: 7727-37-9.
1 DEFINITION
Nitrogen contains NLT 99.0%, by volume, of nitrogen (N₂).
2 IDENTIFICATION
A. The flame of a burning wood splinter is extinguished when inserted into a test tube filled with Nitrogen. [Note-Exercise caution.]
3 ASSAY
3.1 Procedure
Sample: Nitrogen
Standard: Oxygen–helium certified standard (see Reagents, Indicators, and Solutions)
Chromatographic system
- (See Chromatography 〈621〉, System Suitability.)
- Mode: GC
- Detector: Thermal conductivity
- Column: 3-m length × 4-mm inside diameter: molecular sieve prepared from a synthetic alkali-metal aluminosilicate capable of absorbing molecules having diameters of up to 0.5 nm and completely separating oxygen from nitrogen
- Carrier gas: Helium (99.99%)
- Temperature: Thermostatically controlled
Analysis
Samples: Standard and Sample
Introduce the Samples separately into the gas chromatograph by means of a gas sampling valve.
Acceptance criteria: The peak response produced by the Sample exhibits a retention time corresponding to that produced by the Standard and is equivalent to NMT 1.0% of oxygen when compared to the peak response of the Standard, indicating NLT 99.0%, by volume, of N₂.
4 SPECIFIC TESTS
[Note-Reduce the container pressure by means of a regulator. Measure the gases with a gas volume meter downstream from the detector tube to minimize contamination or change of the specimens.]
4.1 Carbon Monoxide
Sample: 1000 ± 50 mL
Analysis: Pass the Sample through a carbon monoxide detector tube (see Reagents, Indicators, and Solutions) at the rate specified for the tube.
Acceptance criteria: NMT 10 ppm
4.2 Limit of Oxygen
Analysis: Determined as directed in the Assay
Acceptance criteria: NMT 1.0%
4.3 Odor
Analysis: Carefully open the container valve to produce a moderate flow of gas. Do not direct the gas stream toward the face, but deflect a portion of the stream toward the nose.
Acceptance criteria: No appreciable odor is discernible.
5 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in cylinders.