Methylphenidate Hydrochloride Tablets
If you find any inaccurate information, please let us know by providing your feedback here

This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Methylphenidate Hydrochloride Tablets contain NLT 93.0% and NMT 107.0% of the labeled amount of Methylphenidate hydrochloride (C₁₄H₁₉NO₂ · HCl).
2 IDENTIFICATION
Change to read:
A. ▲Spectroscopic Identification Tests 〈197〉, Infrared Spectroscopy: 197M▲ (CN 1-May-2020)
Sample: Equivalent to 50 mg of methylphenidate hydrochloride from a portion of powdered Tablets in a 40-mL centrifuge tube. Add 10 mL of
chloroform, shake, and centrifuge. Filter the clear extract through a medium-sized sintered-glass funnel into a beaker, and repeat the
extraction with an additional 10-mL portion of chloroform. Evaporate the combined chloroform extracts on a steam bath to dryness. Agitate
the dried residue with 2 mL of acetonitrile, and filter the mixture through a small sintered-glass funnel. Wash the crystals with an additional
2 mL of acetonitrile, and dry them with the aid of suction.
Acceptance criteria: Meet the requirements
3 ASSAY
Procedure
Buffer: Dissolve 1.6 g of anhydrous sodium acetate in 900 mL of water. Adjust with acetic acid to a pH of 4.0. Dilute with water to 1 L.
Mobile phase: Methanol, acetonitrile, and Buffer (4:3:3)
Internal standard solution: 0.4 mg/mL of Phenylephrine hydrochloride in Mobile phase
Standard stock solution: 0.2 mg/mL of USP Methylphenidate Hydrochloride RS in Mobile phase
Standard solution: Mix 10.0 mL of the Standard stock solution with 5.0 mL of the Internal standard solution
Sample stock solution: 0.2 mg/mL of methylphenidate hydrochloride from finely powdered Tablets (NLT 20 Tablets) prepared as follows.
Dissolve in Mobile phase using 70% of the final volume. Sonicate for 15 min, and cool to room temperature. Dilute with Mobile phase to
volume. Pass a portion of this solution through a suitable membrane filter, discarding the first portion of the filtrate. Avoid the use of glass
filters. Polypropylene filters are suitable for use.
Sample solution: Mix 10.0 mL of the clear filtrate from the Sample stock solution with 5.0 mL of the Internal standard solution.
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 210 nm
Column: 4.6-mm × 25-cm; packing L10
Flow rate: 1.5 mL/min
Injection size: 50 µL
System suitability
Sample: Standard solution
[Note—The relative retention times for phenylephrine hydrochloride and methylphenidate hydrochloride are 0.8 and 1.0, respectively.]
Suitability requirements
Resolution: NLT 2.0 between the analyte and the internal standard peaks
Relative standard deviation: NMT 2.0% from the peak response ratios of the analyte to the internal standard
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of methylphenidate hydrochloride (C₁₄H₁₉NO₂ · HCl) in the portion of Tablets taken:
Result = (Rᵤ / Rₛ) × (Cₛ / Cᵤ) × 100
Rᵤ = peak response ratio of the analyte to the internal standard from the Sample solution
Rₛ = peak response ratio of the analyte to the internal standard from the Standard solution
Cₛ = concentration of USP Methylphenidate Hydrochloride RS in the Standard solution (mg/mL)
Cᵤ = nominal concentration of Methylphenidate Hydrochloride in the Sample solution (mg/mL)
Acceptance criteria: 93.0%–107.0%
4 PERFORMANCE TESTS
Dissolution 〈711〉, Procedure for a Pooled Sample
Medium: Water; 900 mL
Apparatus 1: 100 rpm
Time: 45 min
Analysis: Determine the amount of methylphenidate hydrochloride (C₁₄H₁₉NO₂ · HCl) dissolved by using the procedure in the Assay, making
any necessary volumetric adjustments.
Tolerances: NLT 75% (Q) of the labeled amount of C₁₄H₁₉NO₂ · HCl is dissolved.
Uniformity of Dosage Units 〈905〉: Meet the requirements
5 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in tight containers.
USP Reference Standards 〈11〉
USP Methylphenidate Hydrochloride RS

