Mefloquine Hydrochloride
If you find any inaccurate information, please let us know by providing your feedback here

This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
DOWNLOAD PDF HERE
1 DEFINITION
Mefloquine Hydrochloride contains NLT 98.0% and NMT 102.0% of C17H16F6N2O · HCl, calculated on the anhydrous basis.
2 IDENTIFICATION
Change to read:
A. Spectroscopic Identification Tests 〈197〉, Infrared Spectroscopy: 197K (CN 1-May-2020)
B. Identification Tests—General, Chloride〈191〉
3 ASSAY
Procedure
Solution A: 1.5 g/L of sodium hydrogen sulfate in water
Mobile phase: Dissolve 1 g of tetraheptylammonium bromide in a 1000-mL mixture of acetonitrile, methanol, and Solution A (2:1:2).
System suitability solution: 4 μg/mL each of USP Mefloquine Hydrochloride RS and USP Mefloquine Related Compound A RS in Mobile phase
Standard solution: 0.2 mg/mL of USP Mefloquine Hydrochloride RS in Mobile phase
Sample solution: 0.2 mg/mL of Mefloquine Hydrochloride in Mobile phase
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 280 nm
Guard column: 4-mm × 3-cm; C18 (recommended)
Column: 4.0-mm × 25-cm; 5-μm packing L1
Column temperature: 25°
Flow rate: 0.8 mL/min
Injection size: 20 μL
System suitability
Samples: System suitability solution and Standard solution
[Note—The relative retention times for mefloquine related compound A and mefloquine are about 0.7 and 1.0, respectively.]
Suitability requirements
Resolution: NLT 2.0 between mefloquine related compound A and mefloquine, System suitability solution
Tailing factor: NMT 2.0, Standard solution
Relative standard deviation: NMT 1.0%, Standard solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of mefloquine hydrochloride (C17H16F6N2O · HCl) in the portion of Mefloquine Hydrochloride taken:
Result = (rU/rS) × (CS/CU) × 100
rU = peak response of mefloquine from the Sample solution
rS = peak response of mefloquine from the Standard solution
CS = concentration of USP Mefloquine Hydrochloride RS in the Standard solution (mg/mL)
CU = concentration of Mefloquine Hydrochloride in the Sample solution (mg/mL)
Acceptance criteria: 98.0%–102.0% on the anhydrous basis
4 IMPURITIES
Residue on Ignition 〈281〉: NMT 0.1%
Organic Impurities
Mobile phase: Dissolve 1 g of tetraheptylammonium bromide in a 1-L mixture of a 1.5-g/L solution of sodium hydrogen sulfate, acetonitrile, and methanol (2:2:1).
System suitability solution: 4 μg/mL each of USP Mefloquine Hydrochloride RS and USP Mefloquine Related Compound A RS in Mobile phase. [Note—Mefloquine related compound A is threo-mefloquine.]
Sample stock solution: 4 mg/mL of Mefloquine Hydrochloride in Mobile phase
Sample solution: 4 μg/mL from the Sample stock solution in Mobile phase
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 280 nm
Guard column: 4-mm × 2.5-cm; 5-μm packing L1
Column: 4.0-mm × 25-cm; 5-μm packing L1
Flow rate: 0.8 mL/min
Injection size: 20 μL. [Note—Equilibrate the column with Mobile phase at a flow rate of 0.8 mL/min for 30 min.]
System suitability
Sample: System suitability solution
[Note—The relative retention times for mefloquine related compound A and mefloquine are about 0.7 and 1.0, respectively.]
Suitability requirements
Resolution: NLT 2.0 between mefloquine related compound A and mefloquine
Relative standard deviation: NMT 2.0%
Analysis
Samples: Sample stock solution and Sample solution
Record the chromatogram for a time that is 10 times the retention time of the main peak.
Acceptance criteria: The response of the mefloquine related compound A peak in the Sample stock solution is NMT twice the area of the main peak of the Sample solution (0.2%). The response of any other individual peak, other than the main peak of the Sample stock solution, is NMT that of the main peak of the Sample solution (0.1%); and the sum of the responses of any such peaks of the Sample stock solution is NMT five times the response of the main peak of the Sample solution (0.5%). [Note—Exclude the main peak and any other peak producing a response of less than 0.2 times (0.02%) the main peak of the Sample solution.]
5 SPECIFIC TESTS
Optical Rotation, Specific Rotation〈781S〉
Sample solution: 50 mg/mL in methanol
Acceptance criteria: –0.2° to +0.2°
Water Determination, Method I〈921〉: NMT 3.0%
6 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in tight, light-resistant containers. Store between 15° and 30°.
USP Reference Standards 〈11〉
USP Mefloquine Hydrochloride RS
USP Mefloquine Related Compound A RS threo-Mefloquine.

