Halcinonide Ointment
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Halcinonide Ointment is Halcinonide in a suitable ointment base. It contains NLT 90.0% and NMT 110.0% of the labeled amount of halcinonide (C24H32CIFO5).
2 IDENTIFICATION
A. The retention time of the halcinonide peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
B. The UV spectrum of the halcinonide peak of the Sample solution corresponds to that of the Diluted standard stock solution, as obtained in the test for Organic Impurities.
3 ASSAY
3.1 PROCEDURE
Mobile phase: Acetonitrile and water (50:50)
Internal standard solution: 6 µg/mL of butylparaben in acetonitrile
Standard stock solution: 0.04 mg/mL of USP Halcinonide RS in Internal standard solution
Standard solution: 0.02 mg/mL of USP Halcinonide RS from Standard stock solution prepared as follows. Mix equal volumes of Mobile phase and Standard stock solution.
Sample solution: Nominally 0.02 mg/mL of halcinonide in Mobile phase prepared as follows. Transfer an equivalent to 1 mg of halcinonide from a quantity of Ointment to a glass-stoppered, 50-mL centrifuge tube, and add 25.0 mL of Internal standard solution and 5.0 mL of hexane. Place in a water bath at 58 ± 2° for 3 min, then mix in a vortex mixer for 1 min until the sample is well dispersed. Repeat the above-specified heating and mixing step once. Cool in an ice-methanol bath for 15 min, or until the two phases separate, centrifuging if necessary. Transfer 5.0 mL of the lower layer to a 15-mL centrifuge tube, and add 5.0 mL of Mobile phase.
3.2 Chromatographic system
(See Chromatography (621), System Suitability.)
Mode: LC
Detector: UV 254 nm
Column: 3.9-mm x 30-cm; packing L1
Flow rate: 2 mL/min
Injection volume: 20 µL
System suitability
Sample: Standard solution
[NOTE-The relative retention times for butylparaben and halcinonide are 0.6 and 1.0, respectively.]
Suitability requirements
Resolution: NLT 2.0 between halcinonide and butylparaben
Relative standard deviation: NMT 3.0% for replicate injections
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of halcinonide (C24H32CIFO5) in the portion of Ointment taken:
Result = (RU/RS) x (CS/CU) x 100
RU = peak response ratio of halcinonide to butylparaben from the Sample solution
RS = peak response ratio of halcinonide to butylparaben from the Standard solution
CS = concentration of USP Halcinonide RS in the Standard solution (mg/mL)
CU = nominal concentration of halcinonide in the Sample solution (mg/mL)
Acceptance criteria: 90.0%-110.0%
4 PERFORMANCE TESTS
MINIMUM FILL (755): Meets the requirements
5 IMPURITIES
ORGANIC IMPURITIES
Solution A: 10 mM ammonium acetate in water
Solution B: Acetonitrile
Mobile phase: See Table 1.
Table 1
| Time (min) | Solution A (%) | Solution B (%) |
| 0 | 90 | 10 |
| 20.0 | 10 | 90 |
| 22.0 | 10 | 90 |
| 22.1 | 90 | 10 |
| 25.0 | 90 | 10 |
Diluent 1: Acetonitrile and water (60:30), saturated with hexane
Diluent 2: Acetonitrile and water (10:90)
Standard stock solution: 0.2 mg/mL of USP Halcinonide RS in Diluent 1
Diluted standard stock solution: 0.1 mg/mL of USP Halcinonide RS in Diluent 2 from the Standard stock solution
Standard solution: 0.2 µg/mL each of USP Halcinonide RS in Diluent 2 from the Standard stock solution
Sample stock solution: Nominally 0.2 mg/mL of halcinonide in Diluent 1 prepared as follows. Transfer a portion of Ointment equivalent to 4 mg of halcinonide to a 50-mL centrifuge tube. Add 10 mL of Diluent 1 and 20 mL of hexane heated at 58 ± 2° for 20 min. Shake for NLT 1 min initially to ensure dispersion and at 5-min intervals thereafter. Cool, centrifuge at 3000 rpm for 10 min, and transfer the lower layer to a 20-mL volumetric flask. Repeat the extraction with an additional 5 mL of Diluent 1 and 20 mL of hexane each time. Dilute with Diluent 1 to volume.
Sample solution: 0.1 mg/mL of halcinonide from the Sample stock solution in Diluent 2
Chromatographic system
(See Chromatography (621), System Suitability.)
Mode: LC
Detector: UV 240 nm. For Identification test B, use a diode array detector in the range of 200-450 nm.
Column: 2.1-mm × 15-cm; 1.8-µm packing L1
Flow rate: 0.3 mL/min
Injection volume: 10 µL
System suitability
Sample: Standard solution
[NOTE-See Table 2 for the relative retention times.]
Suitability requirements
Relative standard deviation: NMT 2.4%
Signal-to-noise ratio: NLT 50
Samples: Diluted standard stock solution, Standard solution, and Sample solution
Analysis
[NOTE-The Diluted standard stock solution is used for Identification test B.]
Calculate the percentage of any individual unspecified impurity in the portion of Ointment taken:
Result = (rU/rS) x (CS/CU) x 100
rU = peak response of each unspecified impurity from the Sample solution
rS = peak response of halcinonide from the Standard solution
CS = concentration of USP Halcinonide RS in the Standard solution (µg/mL)
CU = nominal concentration of halcinonide in the Sample solution (µg/mL)
Acceptance criteria: See Table 2.
Table 2
| Name | Relative Retention Time | Acceptance Criteria, NMT (%) |
| Dihydrotriamcinolonea | 0.5 | — |
| Halcinonide | 1.0 | — |
| Any individual unspecified impurity | — | 0.2 |
| Total impurities | — | 3.0 |
a Drug substance process impurity included in the table for identification only.
6 SPECIFIC TESTS
MICROBIAL ENUMERATION TESTS (61) and TESTS FOR SPECIFIED MICROORGANISMS (62): Meets the requirements for the absence of Staphylococcus aureus and Pseudomonas aeruginosa
7 ADDITIONAL REQUIREMENTS
7.1 PACKAGING AND STORAGE
Preserve in well-closed containers. Store at room temperature. Avoid excessive heat.
7.2 USP REFERENCE STANDARDS (11)
USP Halcinonide RS

