Glyburide and Metformin Hydrochloride Tablets
If you find any inaccurate information, please let us know by providing your feedback here

This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Glyburide and Metformin Hydrochloride Tablets contain NLT 90.0% and NMT 110.0% of the labeled amounts of glyburide (C23H28CIN3O5S) and metformin hydrochloride (C4H11N5·HCI).
2 IDENTIFICATION
A. GLYBURIDE: The retention time of the glyburide peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay for Glyburide.
B. Metformin HYDROCHLORIDE: The retention time of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay for Metformin Hydrochloride.
3 ASSAY
3.1 GLYBURIDE
Buffer: 28.8 g/L of monobasic ammonium phosphate
Mobile phase: Acetonitrile and Buffer (40:60). Adjust with 1 N sodium hydroxide to a pH of 5.3.
Diluent: Acetonitrile and water (50:50)
Standard stock solution: 0.25 mg/mL of USP Glyburide RS prepared as follows. Transfer a weighed amount of USP Glyburide RS to a suitable volumetric flask. Dissolve first in the acetonitrile, using 50% of the final volume, and then dilute with water to volume.
Standard solution: 0.025 mg/mL of USP Glyburide RS in Diluent, from the Standard stock solution
System suitability solution 1: Prepare a solution containing 0.025 mg/mL of USP Glyburide Related Compound A RS in Diluent. Transfer 50 µL of this solution to a 50-mL volumetric flask, and dilute with Standard solution to volume.
System suitability solution 2: 5.0 mg/mL of USP Metformin Hydrochloride RS in System suitability solution 1
Sample solution: Dissolve NLT 5 Tablets in Diluent by stirring with a magnetic stirring bar for at least 1 h. Dilute to obtain a solution containing 0.025 mg/mL of glyburide, based on the label claim. Centrifuge a portion of this solution at 3000 rpm for 10 min and use the clear supernatant. [NOTE-Retain a portion of this solution for the Assay for Metformin Hydrochloride.]
Chromatographic system
(See Chromatography (621), System Suitability.)
Mode: LC
Detector: UV 230 nm
Column: 4.6-mm x 15-cm; 5-µm packing L7
Column temperature: 40°
Flow rate: 1.2 mL/min
Injection volume: 100 µL
Run time: 1.25 times the retention time of glyburide
System suitability
Sample: System suitability solution 2
[NOTE-The relative retention time for glyburide related compound A is about 0.30 with respect to glyburide.]
Suitability requirements
Relative standard deviation: NMT 1.5% for the glyburide peak; NMT 10% for the glyburide related compound A peak
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of glyburide (C23H28CIN3O5S) in the portion of Tablets taken:
Result = (rU/rS) x (CS/CU) x 100
rU = peak response of glyburide from the Sample solution
rS = peak response of glyburide from the Standard solution
CS = concentration of USP Glyburide RS in the Standard solution (mg/mL)
CU = nominal concentration of glyburide in the Sample solution (mg/mL)
Acceptance criteria: 90.0%-110.0% of the labeled amount of glyburide
3.2 METFORMIN HYDROCHLORIDE
Buffer: Transfer 1.0 g each of sodium heptanesulfonate and sodium chloride to a 2000-mL volumetric flask. Add 1800 mL of water, and adjust with 0.06 M phosphoric acid to a pH of 3.85. Dilute with water to volume.
Mobile phase: Acetonitrile and Buffer (10:90)
[NOTE-To improve the separation, the composition of acetonitrile and Buffer may be changed to 5:95, if necessary.]
Diluent: Acetonitrile and water (1:40)
Standard solution: 0.25 mg/mL of USP Metformin Hydrochloride RS in Diluent. [NOTE-Sonicate to achieve complete dissolution, if necessary.]
System suitability stock solution: 25 µg/mL each of USP Metformin Related Compound B RS and USP Metformin Related Compound C RS in Diluent
System suitability solution: Transfer 0.5 mL of the System suitability stock solution to a 50-mL volumetric flask, and dilute with Standard solution to volume.
Sample solution: Dilute with water a portion of the retained Sample solution from the Assay for Glyburide to obtain 0.25 mg/mL of metformin hydrochloride based on the label claim.
Chromatographic system
(See Chromatography (621), System Suitability.)
Mode: LC
Detector: UV 218 nm
Column: 3.9-mm x 30-cm; 10-µm packing L1
Column temperature: 30°
Flow rate: 1 mL/min
Injection volume: 5 µL
System suitability
Sample: System suitability solution
[NOTE-The relative retention times for metformin related compound B, metformin, and metformin related compound C are about 0.86, 1.0, and 2.1-2.3, respectively. Metformin related compound C can have a variable retention time. The composition of acetonitrile and Buffer in Mobile phase may be changed to 5:95, if it elutes at a relative retention time of less than 2.1.]
Suitability requirements
Resolution: NLT 1.5 between metformin related compound B and metformin
Tailing factor: 0.8-2.0 for the metformin peak
Relative standard deviation: NMT 1.5% for the metformin peak; NMT 10% each for the metformin related compound B and metformin related compound C peaks
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of metformin hydrochloride (C4H11N5·HCI) in the portion of Tablets taken:
Result = (rU/rS) x (CS/CU) x 100
rU = peak response of metformin from the Sample solution
rS = peak response of metformin from the Standard solution
CS = concentration of USP Metformin Hydrochloride RS in the Standard solution (mg/mL)
CU = nominal concentration of metformin hydrochloride in the Sample solution (mg/mL)
Acceptance criteria: 90.0%-110.0% of the labeled amount of metformin hydrochloride
4 PERFORMANCE TESTS
4.1 DISSOLUTION (711)
4.1.1 Test 1
4.1.1.1 Glyburide
Medium: 0.05 M boric acid and 0.05 M potassium chloride solution. Prepare by dissolving 3.09 g of boric acid and 3.73 g of potassium chloride in 250 mL of water, adjust with 1 N sodium hydroxide to a pH of 9.5, and dilute with water to 1 L; 500 mL.
Apparatus 2: 75 rpm
Time: 30 min
Standard solution: 0.1 mg/mL of USP Glyburide RS prepared as follows. Transfer a weighed amount of USP Glyburide RS to a suitable volumetric flask, dissolve first in acetonitrile, using 20% of the final volume, then dilute with Medium to volume. Dilute further with Medium to obtain a solution having a glyburide concentration, in mg/mL, of (L/500), where L is the label claim of glyburide in mg/Tablet.
Sample solution: Sample per Dissolution (711). Pass a portion of the solution under test through a polypropylene filter of 0.45-µm pore size or a glass fiber filter of 1-µm pore size.
Buffer: 28.8 g/L of monobasic ammonium phosphate in water
Mobile phase: Acetonitrile and Buffer (1:1). Adjust with 1 N sodium hydroxide to a pH of 5.3.
Chromatographic system
(See Chromatography (621), System Suitability.)
Mode: LC
Detector: UV 230 nm
Column: 4.6-mm × 15-cm; 5-µm packing L7
Column temperature: 30°
Flow rate: 1.5 mL/min
Injection volume: 200 µL
System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: 0.8-2.0
Relative standard deviation: NMT 2%
Analysis
Samples: Standard solution and Sample solution
Determine the percentage of the labeled amount of glyburide (C23H28CIN3O5S) dissolved:
Result = (rU/rS) x (CS/L) x V x 100
rU = peak response of glyburide from the Sample solution
rS = peak response of glyburide from the Standard solution
CS = concentration of USP Glyburide RS in the Standard solution (mg/mL)
L = label claim of glyburide (mg/Tablet)
V = volume of Medium, 500 mL
Tolerances: NLT 85% (Q) of the labeled amount of glyburide is dissolved.
4.1.1.2 Metformin hydrochloride
Medium: 0.05 M phosphate buffer, pH 6.8. Prepare by dissolving 6.8 g of monobasic potassium phosphate in 1000 mL of water, and adjust with 0.2 N sodium hydroxide to a pH of 6.8 ± 0.1; 1000 mL.
Apparatus 2: 50 rpm
Time: 30 min
Standard solution: Dissolve a quantity of USP Metformin Hydrochloride RS in Medium, to obtain a solution having a metformin hydrochloride concentration, in mg/mL, of (L/1000), where L is the label claim of metformin hydrochloride in mg/Tablet. Dilute further, if necessary, with Medium.
Sample solution: Sample per Dissolution (711), Pass a portion of the solution under test through a polypropylene filter of 0.45-µm pore size or a glass fiber filter of 1-um pore size. Dilute with Medium, if necessary, to a concentration similar to that of the Standard solution.
Instrumental conditions
(See Ultraviolet-Visible Spectroscopy (857).)
Mode: UV-Vis
Analytical wavelength: 232 nm
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of metformin hydrochloride (C4H11N5·HCI) dissolved:
Result = (AU/AS) x (CS/L) x V x D x 100
AU = absorbance of the Sample solution
AS = absorbance of the Standard solution
CS = concentration of USP Metformin Hydrochloride RS in the Standard solution (mg/mL)
L = label claim (mg/Tablet)
V = volume of Medium, 1000 mL
D = dilution factor for the Sample solution
Tolerances: NLT 85% (Q) of the labeled amount of metformin hydrochloride is dissolved.
4.1.2 Test 2
4.1.2.1 Glyburide
Medium: 0.05 M boric acid and 0.05 M potassium chloride solution. Prepare by dissolving 3.09 g of boric acid and 3.73 g of potassium chloride in 250 mL of water, adjust with 1 N sodium hydroxide to a pH of 9.5, and dilute with water to 1 L; 500 mL.
Apparatus 2: 75 rpm
Time: 30 min
Solution A: Dissolve 0.288 g of sodium lauryl sulfate in 700 mL water.
Mobile phase: Acetonitrile, Solution A, and triethylamine (300:700:2). Adjust with phosphoric acid to a pH of 7.2.
Standard stock solution: 0.1 mg/mL of USP Glyburide RS prepared in acetonitrile
Standard solution: Dilute Standard stock solution with Medium to obtain a solution having a glyburide concentration, in mg/mL, of (L/500), where L is the label claim of glyburide in mg/Tablet.
Sample solution: Sample per Dissolution (711). Pass a portion of the solution under test through a membrane filter of 0.45-µm pore size.
Chromatographic system
(See Chromatography (621), System Suitability.)
Mode: LC
Detector: UV 220 nm
Column: 4.6-mm × 15-cm; 5-µm packing L7
Column temperature: 40°
Flow rate: 2 mL/min
Injection volume
For Tablets labeled to contain 5 mg/500 mg and 2.5 mg/500 mg: 20 µL
For Tablets labeled to contain 1.25 mg/250 mg: 50 µL
Run time: 8 min
System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: NMT 2.0
Relative standard deviation: NMT 2.0%
Analysis
Samples: Standard solution and Sample solution
Determine the percentage of the labeled amount of glyburide (C23H28CIN3O5S) dissolved:
Result = (rU/rS) x (CS/L) x V x 100
rU = peak response of glyburide from the Sample solution
rS = peak response of glyburide from the Standard solution
CS = concentration of USP Glyburide RS in the Standard solution (mg/mL)
L = label claim of glyburide (mg/Tablet)
V = volume of Medium, 500 mL
Tolerances: NLT 80% (Q) of the labeled amount of glyburide is dissolved.
4.1.2.2 Metformin hydrochloride
Medium: 0.05 M phosphate buffer, pH 6.8. Prepare by dissolving 6.8 g of monobasic potassium phosphate in 1000 mL of water, and adjust with 0.2 N sodium hydroxide to a pH of 6.8 ± 0.1; 1000 mL.
Apparatus 2: 50 rpm
Time: 30 min
Solution A: Dissolve 0.288 g of sodium lauryl sulfate in 700 mL of water.
Mobile phase: Acetonitrile, Solution A, and triethylamine (300:700:2). Adjust with phosphoric acid to a pH of 7.2.
Standard solution
For Tablets labeled to contain 5 mg/500 mg and 2.5 mg/500 mg: 0.5 mg/mL of USP Metformin Hydrochloride RS in Medium. Sonicate before final dilution.
For Tablets labeled to contain 1.25 mg/250 mg: 0.25 mg/mL of USP Metformin Hydrochloride RS in Medium. Sonicate before final dilution.
Sample solution: Sample per Dissolution (711), Pass a portion of the solution under test through a membrane filter of 0.45-um pore size.
Chromatographic system
(See Chromatography (621), System Suitability.)
Mode: LC
Detector: UV 254 nm
Column: 4.6-mm × 15-cm; 5-µm packing L7
Column temperature: 40°
Flow rate: 2 mL/min
Injection volume
For Tablets labeled to contain 5 mg/500 mg and 2.5 mg/500 mg: 20 µL
For Tablets labeled to contain 1.25 mg/250 mg: 50 µL
Run time: 8 min
System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: NMT 2.0
Relative standard deviation: NMT 2.0%
Analysis
Samples: Standard solution and Sample solution
Determine the percentage of the labeled amount of metformin hydrochloride (C4H11N5·HCI) dissolved:
Result = (rU/rS) x (CS/L) x V x 100
rU = peak response of the Sample solution
rS = peak response of the Standard solution
CS = concentration of USP Metformin Hydrochloride RS in the Standard solution (mg/mL)
L = label claim (mg/Tablet)
V = volume of Medium, 1000 mL
Tolerances: NLT 85% (Q) of the labeled amount of metformin hydrochloride is dissolved.
4.2 UNIFORMITY OF DOSAGE UNITS (905)
Meet the requirements for Weight Variation for metformin hydrochloride and for Content Uniformity for glyburide
5 IMPURITIES
5.1 GLYBURIDE
Buffer, Mobile phase, Diluent, Sample solution, Chromatographic system, and System suitability: Proceed as directed in the Assay for Glyburide.
Standard solution: Dilute 1.0 mL of the Standard solution from the Assay for Glyburide with Diluent to 100 mL.
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of each glyburide impurity in the portion of Tablets taken:
Result = (rU/rS) x (CS/CU) x (1/F) x 100
rU = peak response of each glyburide impurity from the Sample solution
rS = peak response of glyburide from the Standard solution
CS = concentration of USP Glyburide RS in the Standard solution (mg/mL)
CU = nominal concentration of glyburide in the Sample solution (mg/mL)
F = relative response factor, 1.2 for glyburide related compound A, 1.0 for all other impurities
Acceptance criteria
[NOTE-Disregard any peak less than 0.05%, and disregard any peak observed in the blank.]
Glyburide related compound A: NMT 1.0%
Any other individual impurities: NMT 0.2%
Total impurities: NMT 0.50%, excluding glyburide related compound A
5.2 METFORMIN HYDROCHLORIDE
Buffer, Mobile phase, Sample solution, Chromatographic system, and System suitability: Proceed as directed in the Assay for Metformin Hydrochloride.
Analysis
Sample: Sample solution
Calculate the percentage of each metformin impurity in the portion of Tablets taken:
Result = (rU/rT) x 100
rU = peak response of each metformin impurity from the Sample solution
rT = sum of all the peak responses from the Sample solution
Acceptance criteria
[Note—Disregard any peak less than 0.05%, and disregard any peak observed in the blank.]
Individual metformin impurities: NMT 0.1%
Total impurities: NMT 0.5%
6 ADDITIONAL REQUIREMENTS
6.1 PACKAGING AND STORAGE
Preserve in tight, light-resistant containers, and store at controlled room temperature.
6.2 LABELING
When more than one dissolution test is given, the labeling states the Dissolution test used only if Test 1 is not used.
Change to read:
6.3 USP REFERENCE STANDARDS (11)
USP Glyburide RS
USP Glyburide Related Compound A RS
4-[2-(5-Chloro-2-methoxybenzamido)ethyl]benzenesulfonamide.
C16H17CIN2O4S 368.83(ERR 1-Apr-2021)
USP Metformin Hydrochloride RS
USP Metformin Related Compound B RS
1-Methylbiguanide hydrochloride.
C3H9N5·HCI 151.60
USP Metformin Related Compound C RS
Dimethylmelamine, or N,N-dimethyl-[1,3,5]triazine-2,4,6-triamine.
C5H10N6 154.17

