ETHYLENE OXIDE AND DIOXANE

If you find any inaccurate information, please let us know by providing your feedback here

ETHYLENE OXIDE AND DIOXANE

Ước tính: 1 phút đọc, Ngày đăng:
Cập nhật:

Tóm tắt nội dung

  1. Method I
  2. Method II

This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition

Issued and maintained by the United States Pharmacopeial Convention (USP)

DOWNLOAD PDF HERE

1 Method I

[Caution—Ethylene oxide is toxic and flammable. Prepare these solutions in a well-ventilated fume hood, using great care. Protect both hands and face by wearing polyethylene protective gloves and an appropriate face mask. Store all solutions in hermetic containers, and refrigerate between 4° and 8°.]

[Note—Before using the polyethylene glycol 200 in this test, remove any volatile components from it by placing 500 mL of polyethylene glycol 200 in a 1000-mL round-bottom flask and attaching the flask to a rotary evaporator maintained at 60° and under a vacuum of 10–20mm Hg for 6 h.]

Acetaldehyde solution: 10 μg/mL of acetaldehyde. [Note—Prepare immediately before use.]

Ethylene oxide stock solution: 2.5 mg/g of ethylene oxide.

Prepare as follows: Tare a glass-stoppered conical flask, add 50 mL of polyethylene glycol 200, and reweigh the flask. Transfer 5 mL of the liquid ethylene oxide to a 100-mL beaker chilled in a mixture of sodium chloride and ice (1:3). Transfer 300 μL (corresponding to 250mg) of liquid ethylene oxide to the polyethylene glycol 200, and swirl gently to mix. Replace the stopper, reweigh the flask, and determine the amount of ethylene oxide absorbed by weight difference. Adjust the weight of the mixture with polyethylene glycol 200 to 100.0 g, replace the stopper, and swirl gently to mix. [Note—Fill a chilled pressure bottle with liquid ethylene oxide, and store in a freezer when not in use. Use a small piece of polyethylene film to protect the liquid from contact with the rubber gasket. Use an adequately chilled apparatus where appropriate. Prepare this stock solution immediately prior to use, and store in a refrigerator after preparation.]

Ethylene oxide solution: Tare a glass-stoppered conical flask, and chill it in a refrigerator. Add 35 mL of polyethylene glycol 200, and reweigh the flask. Transfer 1 g of chilled Ethylene oxide stock solution to the tared conical flask. Adjust the weight of the solution with polyethylene glycol 200 to 50.0 g, replace the stopper, and swirl gently to mix. Transfer 10 g of this solution to a 50-mL volumetric flask.

Add 30 mL of water, and mix. Dilute with water to volume, and mix to obtain a solution containing 10 μg/mL of ethylene oxide. [Note—Use an adequately chilled apparatus where appropriate. Prepare immediately before use.]

Dioxane solution: 500 μg/mL of dioxane

Standard solution A: Transfer 0.1 mL of Ethylene oxide solution to a 10-mL pressure headspace vial. [Note—Other sizes such as a 22-mL pressure headspace vial may be used, depending on operating conditions; however, the same size must be used for Standard solution A,

Standard solution B, and the Sample solution.] Add 0.1 mL of Acetaldehyde solution and 0.1 mL of Dioxane solution, seal the vial, and mix.

Standard solution B: Transfer 1.0 g of the test substance to a 10-mL pressure headspace vial, and add 0.1 mL of Ethylene oxide solution, 0.1 mL of Dioxane solution, and 1.0 mL of N,N-dimethylacetamide. Seal the vial, and mix.

Sample solution: Transfer 1.0 g of the test substance to a 10-mL pressure headspace vial, and add 1.0 mL of N,N-dimethylacetamide and 0.2 mL of water. Seal the vial, and mix.

Chromatographic system

(See Chromatography 〈621〉, System Suitability.)

Mode: Headspace GC

Detector: Flame ionization

Column: 0.32-mm × 30-m glass or quartz capillary; 1.0-μm layer of phase G1

Temperature

Injector port: 150°

Detector: 250°

Column: See the column temperature table below.

Initial

Temperature (°)

Temperature Ramp

(°/min)

Final

Temperature (°)

Hold Time at Final

Temperature (min)

50-505
505180-
180302305

Carrier gas: Helium

Linear velocity: 20 cm/s

Injection volume: 1 mL (gaseous headspace)

Injection type: Split ratio 20:1

Headspace sampler

Temperature equilibration time: 45 min

Equilibration temperature

70° for Standard solution A

90° for Standard solution B

90° for Sample solution

Transfer line temperature: 150°

Pressurization time: 1 min

Injection time: 12 s

System suitability

Sample: Standard solution A

[Note—The relative retention times for acetaldehyde and ethylene oxide are 0.94 and 1.0, respectively.]

Suitability requirements

Resolution: NLT 2.0 between acetaldehyde and ethylene oxide

Signal-to-noise ratio: NLT 5, determined from the dioxane peak

Relative standard deviation: NMT 15%

Analysis

Samples: Standard solution B and Sample solution

[Note—The relative retention times for ethylene oxide and dioxane are 1.0 and 2.5, respectively.]

Calculate the content of ethylene oxide, in ppm, in the portion of the test substance taken:

Result = AE × rU/[(rS × WU) − (rU × WS)]

AE = quantity of ethylene oxide added to Standard solution B (μg)

rU = ethylene oxide peak responses from the Sample solution

rS = ethylene oxide peak responses from Standard solution B

WU = weight of the test substance taken to prepare the Sample solution (g)

WS = weight of the test substance taken to prepare Standard solution B (g)

Calculate the content of dioxane, in ppm, in the portion of the test substance taken:

Result = AD × rU/[(rS × WU) − (rU × WS)]

AD = quantity of dioxane added to Standard solution B (μg)

rU = dioxane peak responses from the Sample solution

rS = dioxane peak responses from Standard solution B

WU = weight of the test substance taken to prepare the Sample solution (g)

WS = weight of the test substance taken to prepare Standard solution B (g)

2 Method II

Ethylene oxide standard solution: Dilute 0.5 mL of ethylene oxide in methylene chloride (50 mg/mL)1 with water to 50.0 mL. [Note—The solution is stable for 3 months if stored in vials with polytetra fluoroethylene (polytef)-coated silicon membrane crimped caps at −20°.]

Allow to reach room temperature. Dilute 1.0 mL with water to 250.0 mL to obtain a solution having a concentration of 2 μg/mL of ethylene oxide. [Note—Use this solution immediately after preparation.]

Dioxane standard solution: 0.05 μL/mL of dioxane

Acetaldehyde standard solution: 10 μg/mL of acetaldehyde. [Note—Prepare immediately before use.]

Resolution solution: Add 2.0 mL of Acetaldehyde standard solution and 2.0 mL of Ethylene oxide standard solution to a 10-mL headspace vial. Seal the vial immediately with a polytef-coated silicon membrane and an aluminum cap, and mix carefully.

Standard solution A: 0.48 μg/mL of ethylene oxide, from Ethylene oxide standard solution, and 0.005 μL/mL of dioxane, from Dioxane standard solution, in water

Standard solution B: Transfer 1.0 g of the test substance into a 10-mL headspace vial. Add 2.0 mL of Standard solution A, seal the vial immediately with a polytef-coated silicon membrane and an aluminum cap, and mix carefully.

Sample solution: Transfer 1.0 g of the test substance into a 10-mL headspace vial. Add 2.0 mL of water, seal the vial immediately with a polytef-coated silicon membrane and an aluminum cap, and mix carefully.

Chromatographic system

(See Chromatography 〈621〉, System Suitability.)

Mode: Headspace GC

Detector: Flame ionization

Column: 0.53-mm × 50-m fused-silica capillary column; 5.0-μm layer of phase G27

Temperature

Injector port: 85°

Detector: 250°

Column: See the column temperature table below.

Initial

Temperature (°)

Temperature Ramp

(°/min)

Final

Temperature (°)

Hold Time at Final

Temperature (min)

70102505

Carrier gas: Helium

Flow rate: 4 mL/min

Injection volume: 1 mL (gaseous headspace)

Injection type: Split ratio 3.5 : 1

Headspace sampler

Temperature equilibration time: 30 min

Equilibration temperature: 80°

System suitability

Sample: Resolution solution

[Note—The relative retention times for acetaldehyde and ethylene oxide are 0.9 and 1.0, respectively.]

Suitability requirements

Resolution: NLT 2.0 between acetaldehyde and ethylene oxide

Analysis

Samples: Standard solution B and Sample solution

[Note—The relative retention times for ethylene oxide and dioxane are 1.0 and 1.9, respectively.]

Calculate the content of ethylene oxide, in ppm, in the portion of the test substance taken:

Result = CE × V × rU/[(rS × WU) − (rU × WS)]

CE = concentration of ethylene oxide in the Standard solution A (μg/mL)

V = volume of Standard solution A added to Standard solution B (2.0 mL)

rU = ethylene oxide peak responses from the Sample solution

rS = ethylene oxide peak responses from Standard solution B

WU = weight of the test substance taken to prepare the Sample solution (g)

WS = weight of the test substance taken to prepare Standard solution B (g)

Calculate the content of dioxane, in ppm, in the portion of the test substance taken:

Result = CD × V × ρ × F × rU/[(rS × WU) − (rU × WS)]

CD = concentration of dioxane in Standard solution A (μL/mL)

V = volume of Standard solution A added to Standard solution B (2.0 mL)

ρ = density of dioxane (1.03 g/mL = 1.03 mg/μL)

F = conversion factor (1000 μg/mg)

rU = dioxane peak responses from the Sample solution

rS = dioxane peak responses from Standard solution B

WU = weight of the test substance taken to prepare the Sample solution (g)

WS = weight of the test substance taken to prepare Standard solution B (g)

1 This is a commercially available solution.