Cimetidine
If you find any inaccurate information, please let us know by providing your feedback here

Tóm tắt nội dung
This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
C10H16N6S 252.34
Guanidine, N′′-cyano-N-methyl-N′-[2-[[(5-methyl-1H-imidazol-4-yl)methyl]thio]ethyl]-;
2-Cyano-1-methyl-3-[2-[[(5-methylimidazol-4-yl)methyl]thio]ethyl]guanidine CAS RN®: 51481-61-9; UNII: 80061L1WGD.
1 DEFINITION
Cimetidine contains NLT 98.0% and NMT 102.0% of cimetidine (C10H16N6S), calculated on the dried basis.
2 IDENTIFICATION
Change to read:
A. Spectroscopic Identification Tests 〈197〉, Infrared Spectroscopy: 197K or 197A
Change to read:
B. The UV spectrum of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay. (USP 1- Aug-2022)
3 ASSAY
Change to read:
Procedure
Mobile phase: Methanol, water, and phosphoric acid (20: 80: 0.03)
Standard stock solution: 0.4 mg/mL of USP Cimetidine RS prepared as follows. Accurately weigh and transfer an appropriate amount of USP Cimetidine RS in a suitable flask. Add 20% of the nal volume of methanol to dissolve, and dilute with water to volume. Standard solution: 0.01 mg/mL of USP Cimetidine RS in Mobile phase from the Standard stock solution
Sample stock solution: 0.4 mg/mL of Cimetidine prepared as follows. Accurately weigh and transfer an appropriate amount of Cimetidine in a suitable flask. Add 20% of the nal volume of methanol to dissolve, and dilute with water to volume.
Sample solution: 0.01 mg/mL of Cimetidine in Mobile phase from the Sample stock solution
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 220 nm. For Identification B, use a diode array detector in the range of 190–400 nm. Column: 3.9-mm × 30-cm; packing L1
Flow rate: 2.0 mL/min
Injection volume: 50 µL
System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: NMT 2.0
Relative standard deviation: NMT 2.0%
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of cimetidine (C10H16N6S) in the portion of Cimetidine taken:
Result = (rU/rS) × (CS/CU) × 100
rU = peak response of cimetidine from the Sample solution
rS = peak response of cimetidine from the Standard solution
CS = concentration of USP Cimetidine RS in the Standard solution (mg/mL)
CU = nominal concentration of cimetidine in the Sample solution (mg/mL)
Acceptance criteria: 98.0%–102.0% on the dried basis
4 IMPURITIES
Residue on Ignition 〈281〉: NMT 0.2%
Change to read:
Organic Impurities
Mobile phase: Mix 240 mL of methanol, 0.3 mL of phosphoric acid, 940 mg of sodium 1-hexanesulfonate, and sufficient water to make 1 L, and filter.
Standard solution: 0.80 µg/mL of USP Cimetidine RS in Mobile phase
Sample solution: 0.4 mg/mL of Cimetidine in Mobile phase. Sonicate for NLT 15 min to dissolve. Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 220 nm
Column: 4.6-mm × 25-cm; packing L1
Flow rate: 2.0 mL/min
Injection volume: 50 µL
System suitability
Sample: Standard solution
Suitability requirements
Tailing factor: NMT 2.0
Relative standard deviation: NMT 2.0%
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of each impurity in the portion of Cimetidine taken:
Result = (rU/rS) × (CS/CU) × 100
rU = peak response of each impurity from the Sample solution
rS = peak response of cimetidine from the Standard solution
CS = concentration of USP Cimetidine RS in the Standard solution (μg/mL)
CU = concentration of Cimetidine in the Sample solution (μg/mL)
Acceptance criteria: See Table 1.
Table 1
| Name | Relative Retention Time | Acceptance Criteria, NMT (%) |
| Cimetidine | 1.0 | - |
| Any individual impurity | - | 0.2 |
| Total impurities | - | 1.0 |
5 SPECIFIC TESTS
Delete the following:
Melting Range or Temperature 〈741〉: 139°–144°
Loss on Drying 〈731〉
Analysis: Dry a sample at 110° for 2 h.
Acceptance criteria: NMT 1.0%
6 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in tight, light-resistant containers.
USP Reference Standards 〈11〉
USP Cimetidine RS

