Benzocaine Topical Solution
If you find any inaccurate information, please let us know by providing your feedback here

This article is compiled based on the United States Pharmacopeia (USP) – 2025 Edition
Issued and maintained by the United States Pharmacopeial Convention (USP)
1 DEFINITION
Benzocaine Topical Solution is a solution of Benzocaine in a suitable solvent. It contains NLT 90.0% and NMT 110.0% of the labeled amount of benzocaine (C₉H₁₁NO₂). It contains a suitable antimicrobial agent.
2 IDENTIFICATION
A. The UV spectrum of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
B. The retention time of the major peak of the Sample solution corresponds to that of the Standard solution, as obtained in the Assay.
3 ASSAY
Procedure
Solution A: 0.1% Trifluoroacetic acid, prepared by diluting 1.0 mL of trifluoroacetic acid with water to 1 L
Solution B: Acetonitrile
Mobile phase: See Table 1.
| Time (min) | Solution A (%) | Solution B (%) |
| 0 | 90 | 10 |
| 34 | 50 | 50 |
| 35 | 90 | 10 |
| 38 | 90 | 10 |
Diluent: Solution A and Solution B (1:1)
System suitability solution: 1 µg/mL of USP Benzocaine RS and 2 µg/mL each of USP Aminobenzoic Acid RS and USP Ethyl 4-Nitrobenzoate RS in Diluent
Standard solution: 0.1 mg/mL of USP Benzocaine RS in Diluent
Sample solution: Nominally 0.1 mg/mL of benzocaine in Diluent prepared as follows. Transfer 10 mg of benzocaine from a portion of Topical Solution to a 100-mL volumetric flask, and dilute with Diluent to volume. Pass through a suitable filter of 0.45-µm pore size as needed, discarding the first 2–3 mL of filtrate.
Chromatographic system
(See Chromatography 〈621〉, System Suitability.)
Mode: LC
Detector: UV 280 nm. For Identification test A, use a diode array detector in the range of 200–400 nm.
Column: 4.6-mm × 25-cm; 5-µm packing L7
Flow rate: 1.5 mL/min
Injection volume: 20 µL
System suitability
Samples: System suitability solution and Standard solution
[Note—The relative retention times for aminobenzoic acid, benzocaine, and ethyl 4-nitrobenzoate are 0.3, 1.0, and 2.1, respectively.]
Suitability requirements
Resolution: NLT 6 between aminobenzoic acid and benzocaine; and between benzocaine and ethyl 4-nitrobenzoate, System suitability solution
Tailing factor: NMT 1.5, Standard solution
Relative standard deviation: NMT 1.0%, Standard solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of the labeled amount of benzocaine (C₉H₁₁NO₂) in the portion of Topical Solution taken:
Result = (rᵤ/rₛ) × (Cₛ/Cᵤ) × 100
rᵤ = peak response of benzocaine from the Sample solution
rₛ = peak response of benzocaine from the Standard solution
Cₛ = concentration of USP Benzocaine RS in the Standard solution (mg/mL)
Cᵤ = nominal concentration of benzocaine in the Sample solution (mg/mL)
Acceptance criteria: 90.0%–110.0%
4 IMPURITIES
Organic Impurities
Solution A: 0.1% Trifluoroacetic acid, prepared by diluting 1.0 mL of trifluoroacetic acid with water to 1 L
Solution B: Acetonitrile
Mobile phase: See Table 2.
| Time (min) | Solution A (%) | Solution B (%) |
| 0 | 85 | 15 |
| 34 | 55 | 45 |
| 35 | 85 | 15 |
| 38 | 85 | 15 |
Diluent: Solution A and Solution B (1:1)
Standard solution: 1 µg/mL of USP Benzocaine RS and 2 µg/mL each USP Aminobenzoic Acid RS and USP Ethyl 4-Nitrobenzoate RS in Diluent
Sample solution: Nominally 1 mg/mL of benzocaine in Diluent prepared as follows. Transfer 50 mg of benzocaine from a portion of Topical Solution to a volumetric flask and dissolve with aid of sonication as needed, then dilute with Diluent to volume. Pass through a suitable filter of 0.45-µm pore size as needed, discarding the first 2–3 mL of filtrate.
Chromatographic system
Mode: LC
Detector: UV 280 nm
Column: 4.6-mm × 25-cm; 5-µm packing L7
Flow rate: 1.5 mL/min
Injection volume: 20 µL
System suitability
Sample: Standard solution
[Note—See Table 3 for relative retention times.]
Suitability requirements
Resolution: NLT 6 between aminobenzoic acid and benzocaine; and between benzocaine and ethyl 4-nitrobenzoate
Relative standard deviation: NMT 2.0% for each peak corresponding to benzocaine, aminobenzoic acid, and ethyl 4-nitrobenzoate
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of aminobenzoic acid and ethyl 4-nitrobenzoate in the portion of Topical Solution taken:
Result = (rᵤ/rₛ) × (Cₛ/Cᵤ) × 100
rᵤ = peak response of aminobenzoic acid or ethyl 4-nitrobenzoate from the Sample solution
rₛ = peak response of aminobenzoic acid or ethyl 4-nitrobenzoate from the Standard solution
Cₛ = concentration of USP Aminobenzoic Acid RS or USP Ethyl 4-Nitrobenzoate RS in the Standard solution (mg/mL)
Cᵤ = nominal concentration of benzocaine in the Sample solution (mg/mL)
Calculate the percentage of any other individual unspecified impurity in the portion of Topical Solution taken:
Result = (rᵤ/rₛ) × (Cₛ/Cᵤ) × 100
rᵤ = peak response of any other individual unspecified impurity from the Sample solution
rₛ = peak response of benzocaine from the Standard solution
Cₛ = concentration of USP Benzocaine RS in the Standard solution (mg/mL)
Cᵤ = nominal concentration of benzocaine in the Sample solution (mg/mL)
Acceptance criteria: See Table 3. Disregard peaks less than 0.05%.
| Name | Relative Retention Time | Acceptance Criteria, NMT (%) |
| Aminobenzoic acid | 0.27 | 0.20 |
| Benzocaine | 1.0 | — |
| Ethyl 4-nitrobenzoate | 2.5 | 0.20 |
| Any other individual unspecified impurity | — | 0.10 |
| Total impurities | — | 1.0 |
5 SPECIFIC TESTS
Microbial Enumeration Tests 〈61〉 and Tests for Specified Microorganisms 〈62〉: It meets the requirements of the tests for absence of Staphylococcus aureus and Pseudomonas aeruginosa.
6 ADDITIONAL REQUIREMENTS
Packaging and Storage: Preserve in tight containers, protected from light, and avoid prolonged exposure to temperatures exceeding 30°.
USP Reference Standards 〈11〉
USP Aminobenzoic Acid RS
Benzoic acid, 4-amino. C₇H₇NO₂ 137.14
USP Benzocaine RS
USP Ethyl 4-Nitrobenzoate RS
Benzoic acid, 4-nitro-, ethyl ester. C₉H₉NO₄ 195.17

